临床试验/NCT02267577
NCT02267577已完成不适用

Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Adult Volunteers

William Marsh Rice University1 个研究点 分布在 1 个国家目标入组 90 人开始时间: 2015年1月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
90
试验地点
1
主要终点
Percentage of Sphygmo Readings That Were With < 5 mmHg, < 10 mmHg, and < 15 mmHg of the Readings by the Gold Standard Sphygmomanometer.

研究概览

简要总结

The research team has developed an automatic blood pressure monitor (Sphygmo) to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited.

90 adult volunteers will be enrolled by researchers at Rice University. The participant will be seated in a comfortable chair with arm at heart level. Arm circumference will be measured and a blood pressure cuff will be placed on the arm. The cuff will be inflated and blood pressure measurements will be taken by a commercially available device and by the Sphygmo device. Blood pressure measurements from both devices will be recorded. The participant's blood pressure will be measured up to 9 times with a waiting period of 45-60 seconds between each measurement.

The results of this study will be used to optimize the blood pressure detection algorithm and thus further develop the device.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Device Feasibility
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者
是

入选标准

  • •18 years of age or older
  • •Willing and able to provide informed consent

排除标准

  • •Unable to provide informed consent

研究组 & 干预措施

Healthy Adults Volunteers

Experimental

Men and women over the age of 18 will have their blood pressure measured with both the Sphygmo: Automatic Blood Pressure Monitor device and the GE Dinamap ProCare automatic blood pressure monitor.

干预措施: Sphygmo: Automatic Blood Pressure Monitor (Device)

Healthy Adults Volunteers

Experimental

Men and women over the age of 18 will have their blood pressure measured with both the Sphygmo: Automatic Blood Pressure Monitor device and the GE Dinamap ProCare automatic blood pressure monitor.

干预措施: GE Dinamap ProCare Automatic Blood Pressure Monitor (Device)

结局指标

主要结局

Percentage of Sphygmo Readings That Were With < 5 mmHg, < 10 mmHg, and < 15 mmHg of the Readings by the Gold Standard Sphygmomanometer.

时间窗: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.

The British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of \< 5 mmHg, \< 10 mmHg, and \< 15 mmHg. A letter grade of "A" requires that over 60%, 85%, and 95% were achieved in the \< 5 mmHg, \< 10 mmHg, and \<15 mmHg categories, respectively. A letter grade of "B" requires that over 50%, 75%, and 90% were achieved in the \< 5 mmHg, \< 10 mmHg, and \<15 mmHg categories, respectively. The test device achieved a grade of (B/B) with the validation data from this study.

Accuracy of Blood Pressure Measurement by Sphygmo Relative to the Clinical Standard.

时间窗: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.

Mean difference of systolic and diastolic blood pressure readings between Sphygmo device and the gold standard sphygmomanometer.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Rebecca Richards-Kortum

Stanley C. Moore Professor of Bioengineering

William Marsh Rice University