SpO2 Validation Study - Philips Optimus

Philips Clinical & Medical Affairs Global1 个研究点 分布在 1 个国家目标入组 24 人开始时间: 2025年9月9日最近更新:
干预措施

试验速览

阶段
不适用
状态
终止
入组人数
24
试验地点
1
主要终点
SpO2 accuracy of the SpO2 sensors under test with the Philips MRI Patient Monitoring

研究概览

简要总结

Validation study to validate the SpO2 accuracy performance of the Sponsor's SpO2 sensors over the range of 70-100% SaO2 during non-motion conditions assessed by CO-Oximetry

详细描述

The intent of this clinical study is to validate the SpO2 accuracy performance of the Sponsor's SpO2 sensors over the range of 70-100% SaO2 during non-motion conditions assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the Sponsor's pulse oximeter with consideration to anticipated updates to ISO 80601-2-61:2025(E)

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Participants accepted in this study must meet all the following criteria:
  • Participant must have the ability to understand and provide written informed consent.
  • Participant is an adult between 18-50 years of age.
  • Participant must be willing and able to comply with study procedures and duration.
  • Participant is a non-smoker or who has not smoked within 2 days prior to the study.

排除标准

  • Participant is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation (i.e., Raynaud's Syndrome), injury, open wound, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
  • Females who are pregnant
  • Females who are trying to get pregnant (with confirmation of positive urine pregnancy test unless the participant is known to be not of child-bearing potential)
  • Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
  • Participants with known respiratory conditions such as: (self-reported)
  • uncontrolled / severe asthma,
  • pneumonia / bronchitis,
  • shortness of breath / respiratory distress,
  • unresolved respiratory or lung surgery,
  • emphysema, COPD, lung disease
  • Recent COVID with or without hospitalization (last 2 months)
  • Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)
  • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg on 3 consecutive readings (reviewed during health screen)
  • have had cardiovascular surgery, except successful minor surgery without clinical symptoms (i.e., PFO, PDA)
  • chest pain (angina)
  • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
  • previous heart attack
  • blocked artery
  • unexplained shortness of breath
  • congestive heart failure (CHF)
  • history of stroke
  • transient ischemic attack
  • carotid artery disease
  • myocardial ischemia
  • myocardial infarction
  • cardiomyopathy
  • implantable active medical device such as pacemaker or automatic defibrillator
  • Self-reported health conditions as identified in the Health Assessment Form
  • diabetes,
  • uncontrolled thyroid disease,
  • kidney disease / chronic renal impairment,
  • history of seizures (except childhood febrile seizures),
  • epilepsy,
  • history of unexplained syncope,
  • recent history of frequent migraine headaches,
  • recent symptomatic head injury (within the last 2 months),
  • Cancer requiring chemotherapy, radiation, or current treatment.
  • Participants with known clotting disorders (self-reported)
  • history of bleeding disorders or personal history of prolonged bleeding from injury
  • history of blood clots
  • hemophilia
  • current use of blood thinner: prescription or daily use of aspirin
  • Sickle Cell Trait or Disease
  • Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported)
  • Participants with severe allergies to iodine (only applicable if iodine is used)
  • Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine),
  • Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
  • Unwillingness or inability to remove nail polish or artificial nails from test digits.
  • Participant received intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
  • 另有 3 项未显示

研究组 & 干预措施

Philips Wireless Patient Module SpO2 (Gen-4)

This device is a complete pulse oximeter measurement module that utilizes the Philips PicoSAT II+ SpO2 processor that is a component on the circuit car assembly, part number 453665151651. This processor incorporates all signal processing and algorithms to derive the oxygen saturation value (SpO2), pulse rate, perfusion indicator value, and the plethysmographic waveform

干预措施: No intervention (Device)

SpO2 Sensor, MRI (Gen-4)

This is a fiber-optic cable that is attached via an connector at one end to the Philips Wireless SpO2 Patient Module, and one of the SpO2 finger accessories at the other end. This cable serves as the conduit through which red and infrared light is transmitted into the finger and then is received back from the other side of the finger into the photodiode. The red and infrared LEDs and the photodiode are all contained in the cable housing. The cable itself is made from RoHS-compliant fiber-optic materials. Fiber-optics are utilized to allow the SpO2 module to operate properly in the MRI environment.

干预措施: No intervention (Device)

SpO2 Accessories: SpO2 Grips (Single Use) and SpO2 Gloves (Reusable)

The SpO2 Grips and Gloves are designed to securely hold the SpO2 Sensor in proper location when it is attached to the Wireless SpO2 module. Both finger accessories provide two slide-in clip attachments for connecting to the SpO2 Cable ends, designed to be positioned at the top and bottom of the finger. The reusable finger accessory is designed to slide onto the finger and secure itself by means of a rubber-like material that provides the appropriate grip

干预措施: No intervention (Device)

Silver Reference: IntelliVue Patient Monitor X3 and Adult SpO2 Sensor (M1191BL)

The IntelliVue X3 is a versatile patient monitoring device with a color touchscreen display. It can simultaneously monitor ECG (using 3-, 5-, 6- or 10-lead sets, including arrhythmia and ST monitoring), respiration, SpO2, NBP, two invasive pressures, temperature, and CO2. The multi-touch screen allows easy interaction by swiping and tapping with one or two fingers.

干预措施: No intervention (Device)

结局指标

主要结局

SpO2 accuracy of the SpO2 sensors under test with the Philips MRI Patient Monitoring

时间窗: Day 1

Pulse Oximetry Subsystem during non-motion conditions over the range of 70-100% SaO2, when compared with the reference SaO2 obtained from CO-Oximetry analysis of arterial blood samples.

次要结局

  • evaluate non-disparate performance (differential bias) of the SpO2 sensors under test(Day 1)

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...

相似试验