Agreement Between Standard and Continuous Wireless Vital Sign Measurements Following Major Abdominal Surgery: A Clinical Comparison Study

Rigshospitalet, Denmark1 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2019年12月18日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
20
试验地点
1
主要终点
Heart rate comparisons

研究概览

简要总结

Patients undergoing major surgery are at risk of postoperative complications. Continuous wireless monitoring outside the post-anesthesia or intensive care units may enable early detection of patient deterioration, but good accuracy of measurements is required.

This validation study, which is part of the WARD-SX project, aimed to assess the agreement between vital signs recorded by standard and novel wireless devices

详细描述

Patients undergoing major surgery are at risk of postoperative complications. Complications may be preceded by abnormal vital signs. Standard monitoring at general wards is intermittent and manual, leaving the patients unobserved for extended periods.

The chance to enhance monitoring has arisen with recent technological advances in the form of small wearable and wireless devices that continuously record vital signs. Wearable devices enable continuous monitoring outside the intensive care unit and post-anesthesia care unit, thus enabling early detection of patient deterioration. However, before implementing such devices in hospital settings, good accuracy, and precision of measurements is required.

This validation study, which is part of the WARD-SX project, aimed to assess the agreement between vital signs recorded by standard and novel wireless devices

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
60 Years 至 110 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • ≥60 years, scheduled for major abdominal cancer surgery with planned PACU admission.
  • Estimated surgical intervention duration ≥2 hours.

排除标准

  • Implanted cardioverter defibrillator or pacemaker, allergy to study devices, severe cognitive impairment assessed by Mini-Mental State Examination ≤24, or inability to cooperate in wearing the wireless monitoring equipment.

结局指标

主要结局

Heart rate comparisons

时间窗: 1.5 hours (measurement interval every 15 min)

Heart rate (beats per minute) measurements compared between Isansys Lifetouch and standard monitoring (Phillips IntelliVue). Analysed using Bland Altman (BA) analysis to measure mean diference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

SpO2 measurements comparisons wireless vs. invasive

时间窗: 1.5 hours

SpO2 (percent) measurements compared between Nonin WristOx 3150 and arterial arterial oxygen saturation. Analysed using Bland Altman (BA) analysis to measure mean difference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

Blood pressure comparisons, wireless vs. arterial BP

时间窗: 1.5 hours

Systolic and diastolic (mmHg) measurements compared between Meditech BlueBp-05 and invasive arterial blood pressure. Analysed using Bland Altman (BA) analysis to measure mean difference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

Respiratory rate comparisons

时间窗: 1.5 hours

Respiratory rate (breaths per minute) measurements compared between Isansys Lifetouch and observed counted measurements. Analysed using Bland Altman (BA) analysis to measure mean difference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

SpO2 measurements comparisons wireless vs. standard

时间窗: 1.5 hours (measurement interval every 15 min)

SpO2 (percent) measurements compared between Nonin WristOx 3150 and standard monitoring. Analysed using Bland Altman (BA) analysis to measure mean difference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

Pulse rate comparisons

时间窗: 1.5 hours

Pulse rate measurements (beats per minute) measures compared between Nonin WristOx 3150 and standard monitoring. Analysed using Bland Altman (BA) analysis to measure mean difference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

Blood pressure comparisons wireless vs. standard

时间窗: 1.5 hours, every 15 min

Blood pressure (mmHg) measurements compared between Meditech BlueBp-05 and standard monitoring. Analysed using Bland Altman (BA) analysis to measure mean difference and limits of agreement between to methods. The BA analysis is corrected for repeated measures within each subject.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Camilla Haahr-Raunkjaer

MD.

Rigshospitalet, Denmark