Clinical Evaluation of CARESTREAM Cone Beam Computed Tomography (CBCT)
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 48
- 主要终点
- Radlex Scale for Diagnostic Quality Ratings - 2D Images
研究概览
简要总结
The purpose of this study is to evaluate the 2D and 3D imaging performance of the CARESTREAM Cone Beam Computed Tomography (CBCT) scanner ("investigational device") against the commercially available 2D and 3D predicate devices. The results of this study will be included in a Traditional 510(k) FDA Submission to obtain clearance to market the new CARESTREAM Cone Beam Computed Tomography (CBCT) scanner in the US. The study was designed in accordance with the FDA Guidance titled "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices", issued on August 6, 1999.
详细描述
The purpose of this study is to evaluate the 2D and 3D imaging performance of the CARESTREAM Cone Beam Computed Tomography (CBCT) scanner ("investigational device") against the commercially available 2D and 3D predicate devices. The study undertaken will demonstrate the safety and effectiveness of the investigational CARESTREAM Cone Beam Computed Tomography (CBCT) 2D imaging functionality by comparing against the predicate CARESTREAM DRX-1 GOS detector used on adult cadaver specimens to capture the image pairs.
The study will demonstrate the safety and effectiveness of the investigational CARESTREAM Cone Beam Computed Tomography (CBCT) 3D imaging functionality by comparing against the reference PHILLIPS Multi Detector Computed Tomography (MDCT) scanner used on adult cadaver specimens (body part) to capture 3D volumetric exams using the anatomy of both upper and lower extremity specimens. The adult cadaver specimens will receive multiple x-ray exposures using the predicate devices for both 2D and 3D imaging. Target images from cadaver specimens will be reviewed by radiologist readers to rate diagnostic image quality, and the data will be included in a comparative evaluation.
The live human subject portion of the study will be performed on healthy volunteers. Each volunteer will sign an Informed Consent after which will receive only one 3D volumetric exam using the investigational CARESTREAM Cone Beam Computed Tomography (CBCT) scanner. Target images will be reviewed by radiologist readers to rate diagnostic image quality according to a radiologist reader scale applicable to 3D volumetric imaging.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Diagnostic
- 盲法
- Single (Outcomes Assessor)
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Subject 18 years or older
- •Subject has provided informed consent
- •Subject is able to be positioned properly in the investigational device and be still during the exam to reduce the potential of motion in the images.
- •Subject(s) may have a metal screw, plate or artificial joint
排除标准
- •Subject is pregnant
- •Not able or willing to provide Informed Consent, or consent is withdrawn
- •Not able to collect all required case information
- •Subject has a history of high radiation exposure:
- •Has undergone radiation therapy
结局指标
主要结局
Radlex Scale for Diagnostic Quality Ratings - 2D Images
时间窗: 12 weeks after last image capture
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - 3D Images SND
时间窗: 12 weeks after last image capture
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - 3D Images FDK
时间窗: 12 weeks after last image capture
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - 3D Images High Resolution
时间窗: 12 weeks after last image capture
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
次要结局
未报告次要终点
