临床试验/NCT04384094
NCT04384094Unknown不适用

Rebound Esthesiometer Phase 2 Protocol

Icare Finland Oy0 个研究点目标入组 100 人开始时间: 2020年8月最近更新:
适应症

试验速览

阶段
不适用
入组人数
100
主要终点
Rebound esthesiometer operating parameters

研究概览

简要总结

The purpose of this study is to define the operating parameters for a new method to measure corneal sensitivity.

详细描述

Corneal sensitivity is the most important protective mechanism of the eye. Thus, measuring it at any given time gives important indicators of corneal physiology, especially in the diagnostics of corneal and systemic diseases (e.g. diabetes, herpes simplex and keratitis) and recovery from ocular surgery. Current commercial measurement methods are outdated, non-quantitative and uncomfortable to use, hence left unused. This results in suboptimal, even poor, diagnosis and treatment for patients.

In this study the operating parameters for a state-of-the-art esthesiometer are defined. The operating principle is based on existing rebound technology which is already approved for IOP measurement.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age >= 18
  • Informed consent signed

排除标准

  • Signs of infectious or inflammatory disease of anterior eye at the time of presentation.
  • During last two weeks, any of the following events or conditions occurred or was present: inflammatory or infectious ocular condition, surgical or other intervention or therapy, other abnormal ocular event.
  • Unable to give informed consent.
  • Directly or indirectly indicated vunerability.

结局指标

主要结局

Rebound esthesiometer operating parameters

时间窗: 2020-2021

Measure the correlation of corneal touch sensitivity between rebound esthesiometer and the reference instrument Cochet-Bonnet monofilament esthesiometer. With the rebound esthesiometer the impact is varied by changing the speed (m/s) of the measurement probe. The probe mass (mg) and speed (m/s) equals to the touch impact impulse which causes the touch sensation. The subjective sensation of the test subject is recorded.

Rebound esthesiometer operating threshold

时间窗: 2020-2021

Define the corneal touch sensitivity threshold using the rebound esthesiometer by varying the measurement probe impact at the moment of touch of corneal surface. The impact is varied by changing the speed (m/s) of the measurement probe and the subjective sensation of the test subject is recorded. The mass (mg) of the measurement probe is known and the impact energy is calculated.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor