Accuracy of Real Time 3D Navigation Device to Identify Lumbar Epidural Spacy in Pregnant Patients. A Comparison With Standard Ultrasound.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 90
- 试验地点
- 1
- 主要终点
- epidural space depth
研究概览
简要总结
The aim of this double blinded study is to assess the abilities of the SpineNav3DTM technology (Accuro) to identify the lumbar epidural space in parturients using as comparator a standar ultrasound (US) device.
详细描述
SpineNav3DTM technology facilitates image interpretation of individual 2D lumbar spine scans by automating spinal bone landmark detection and depth measurements and providing a real-time assessment of scan plane orientation in 3D.
Assuming a Type I error rate of 5% (α = 0.05) and a Type II error rate of 20% (β=0.2), a minimum of 76 subjects will be required. All the subjects will have their lumbar area scanned for the measurements of the depth of the epidural space (recorded in cm) by two anesthesiologist, one expert in UD and the other a novice, previously instructed to the use of the Accuro device.
All measurements will be performed either in the sitting or in the supine position at L1/L2, L2/L3 and L3/L4 levels. The expert operator will assess the lumbar spine by using both the devices (standard US and Accuro) while the novice will use only the Accuro.
The primary end-point of the study will be the accuracy of the Accuro to determine the depth of the epidural space using a midline approach using as comparator the measurement obtained with the standard ultrasound device performed by the expert operator.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 45 Years(Adult)
- 性别
- Female
- 接受健康志愿者
- 是
入选标准
- •Healthy pregnant women undergoing antenatal anesthetic consultation
排除标准
- 未提供
结局指标
主要结局
epidural space depth
时间窗: day 1
measurement of the depth of the epidural space
次要结局
未报告次要终点
研究者
Giorgio Capogna
Head Department of Anesthesia
Città di Roma Hospital