临床试验/NCT05998629
NCT05998629招募中不适用

Skin-interfaced Colorimetric Bifluidic Sweat Sensor Device for the Diagnosis of Cystic Fibrosis (CF)

Milton S. Hershey Medical Center1 个研究点 分布在 1 个国家目标入组 30 人开始时间: 2024年7月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
招募中
入组人数
30
试验地点
1
主要终点
Sweat chloride concentration using standard of care lab assessment

研究概览

简要总结

Cystic fibrosis (CF) is a multisystem autosomal recessive inherited disease affecting approximately 75,000 individuals in USA. The sweat chloride (Cl) test remains the gold standard for diagnosis of CF but still has a number of limitations. The objectives of this study are: 1)To evaluate a skin-interfaced colorimetric bifluidic sweat device with two synchronous channels as a potential low-cost but potentially accurate test to diagnoses cystic fibrosis (CF) and 2) To evaluate measurements of sweat chloride (Cl) using this same system in comparison to the standard clinical laboratory procedures routinely performed in the Clinical Laboratory at Penn State Health Milton S. Hershey Medical Center (PSH-HMC), Hershey, PA for assessment of the diagnosis of CF. This is a single institution study performed solely at PSH-HMC. Study participants will include 1) adults 18 years of age or older capable of providing signed and dated informed consent, 2) subjects with an established known diagnosis of cystic fibrosis (CF) or healthy volunteers, and 3) able to understand and speak English language. Exclusion criteria include: 1) any medical condition or disorder known to potentially interfere with accurate measurements of sweat chloride and 2) inability to understand and speak the English language. Cystic Fibrosis (CF) subjects will be identified from the population of eligible patients receiving medical care at Penn State Health- Milton S. Hershey Medical Center (PSH-HMC). Healthy donor volunteers will be recruited from various members of the PSH-HMC CF clinical care team, members of the Division of Allergy, Pulmonary and Critical Care (both faculty and trainees) at PSH-HMC, and PSU-University Park research team. The total projected number of combined enrolled subjects is 30. This is a single day single time study that will require approximately 60 minutes of subject participation. Potential risks include a) side effects from pilocarpine iontophoresis sweat test collection (pain, skin discomfort, blisters, rarely burns and b) loss of confidentiality. There will be no cost to subjects for study participation. There will be no reimbursement financially for study participation. There is no benefit to subjects for study participation. There is the potential benefit to medical science via identification of improved method to accurately measure sweat chloride for diagnosis of CF.

详细描述

Background: Cystic fibrosis (CF) is a multisystem autosomal recessive inherited disease affecting approximately 75,000 individuals in USA. The sweat chloride (Cl) test remains the gold standard for diagnosis of CF but still has a number of limitations including 1) common inability to recover enough sweat volume for test accuracy, termed "quantity not sufficient" limitation, 2) difficulty of performing in young children and infants, 3) absence of point of care (POC) simplicity and need to schedule prolonged (> 1 hour) appointment with high resource and personnel utilization, and 4) still a range of values without one specific diagnostic cut-off measurement thus creating sweat Cl classifications of negative (normal), indeterminant, and positive (diagnostic). Multiple publications already exist validating the use of POC skin-interfaced micro fluidic colorimetric systems in attempts to overcome these limitations of the still standard of care macroduct sweat collection and analysis system. The current proposed investigational diagnostic (not therapeutic) device introduces a skin-interfaced colorimetric bifluidic sweat collection and analyses device with two synchronous channels to quantify sweat rate and biomarkers (for this study Cl) in real-time, even during uncertain sweat conditions and activities. This proposed bifluidic system could provide accurate analysis of biomarkers (including Cl) based upon instantaneous sweat rate. For consistent of data accrual and analyses in this proposed study, for both methods of sweat collection pilocarpine iontophoresis will be the method of sweat stimulation. This skin-interfaced colorimetric bifluidic sweat device with two synchronous channels device is a closed system design to reduce evaporation and resists contamination from the external environment. This low-cost yet highly accurate device provides opportunities for clinical sweat analysis and disease screening in remote and low-resource settings which is currently non-existent with standard clinical measurements of sweat Cl. In addition, this device can be easily adapted for other biomarkers, when corresponding colorimetric reagents are exploited.

Inclusion Criteria: 1) Adults 18 years of age or older capable of providing signed and dated informed consent, 2) Subjects with an established know diagnosis of cystic fibrosis (CF) or healthy volunteers, 3) Competent to understand written and spoken English language.

Exclusion Criteria: Participants receiving medications or with known disorders that can cause errors in sweat test analysis. Common causes of potential error inducing factors in sweat test determination include Lithium therapy, mineralocorticoid hormone therapy, adrenal insufficiency, glycogen storage diseases, hypothyroidism, hypoparathyroidism, nephrogenic diabetes insipidus, G6PD deficiency, ectodermal dysplasia, or any skin or soft tissue disorders that could affect obtaining the necessary volume and quality of sweat. Non-English-speaking subjects will be excluded.

Study Design: This is a single institution study to be performed entirely at the Penn State Health Milton S. Hershey Medical Center (PSH-HMC), Hershey, PA. This is only a single day, single time study, i.e. VISIT 1 and DAY 1 ONLY. The study involves minimal risk. There is no anticipated benefit to subjects for study participation. Total subject participation time will be approximately 60 minutes, after which subjects' participation will end. The total combined estimated enrollment for both health volunteer subjects and CF subjects will be 30. Two sweat chloride samples and measurements as per study procedures will be obtained simultaneously from both arms using pilocarpine iontophoresis.

Study Procedures: 1) PSH-HMC Clinical Laboratory Department of Pathology SOP dated 14-Nov-2022 "Macroduct Sweat Collection" will be complied for attainment of the sample for which sweat chloride will be measured at PSH-HMC. The site for this procedure is Suite 520 at the University Physicians Center, Hershey, PA. After completion of pilocarpine iontophoresis and following collection of sweat specimens inside the macroduct tubing, the sweat will then be eluted and analyzed separately. PSH-HMC sample chloride biochemical analyses: a) add reached 12 μL as indicated in the microfluidic device. The measured color change distance will be used to determine the chloride concentration according to the linear relationship.2N nitric acid and s-Diphenylcarbazone, b) rinse with 0.005N mercuric nitrate and titrate to very light purple color and record volume of mercuric nitrate, c) prepare control standards, d) calculate the chloride result using the EXCEL calculation template.

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Parallel
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 80 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • adults 18 years of age or older capable of providing signed and dated informed consent,
  • subjects with an established known diagnosis of cystic fibrosis (CF) or healthy volunteers,
  • able to understand and speak English language.

排除标准

  • any medical condition or disorder known to potentially interfere with accurate measurements of sweat chloride
  • inability to understand and speak the English language.

研究组 & 干预措施

Healthy control subjects experimental device

Experimental

a skin-interfaced colorimetric bifluidic sweat device with two synchronous channels for healthy control subjects

干预措施: a skin-interfaced colorimetric bifluidic sweat device with two synchronous channels (Device)

Healthy control subjects standard of care

Active Comparator

standard clinical laboratory procedures routinely performed in the Clinical Laboratory at Penn State Health Milton S. Hershey Medical Center (PSH-HMC), Hershey, PA for measurement of sweat chloride concentrations for healthy control subjects

干预措施: Standard of Care laboratory procedure for measurement of sweat (Other)

Cystic Fibrosis Subjects experimental device

Experimental

a skin-interfaced colorimetric bifluidic sweat device with two synchronous channels for cystic fibrosis subjects

干预措施: a skin-interfaced colorimetric bifluidic sweat device with two synchronous channels (Device)

Cystic Fibrosis Subjects standard of care

Active Comparator

standard clinical laboratory procedures routinely performed in the Clinical Laboratory at Penn State Health Milton S. Hershey Medical Center (PSH-HMC), Hershey, PA for measurement of sweat chloride concentrations for cystic fibrosis subjects

干预措施: Standard of Care laboratory procedure for measurement of sweat (Other)

结局指标

主要结局

Sweat chloride concentration using standard of care lab assessment

时间窗: Single point in time measurement 60 minutes after device attachment

Sweat chloride concentration in mmol/L

次要结局

  • Sweat chloride concertation using skin-interfaced colorimetric device(Single point in time measurement 60 minutes after device attachment)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Robert Vender

Principal Investigator

Milton S. Hershey Medical Center

研究点 (1)

Loading locations...