A Feasibility Study Into the Implant of the WiSE CRT System With an Intracardiac Pacemaker to Achieve Totally Leadless CRT
试验速览
- 阶段
- 不适用
- 状态
- 尚未招募
- 入组人数
- 40
- 主要终点
- Safety from major complications post device implants & procedures
研究概览
简要总结
This study is a single arm, prospective, feasibility, multi-centre, observational study. Participants will be suitable for a Cardiac Resynchronisation Therapy (CRT) implant using Wireless Stimulation Endocardially for Cardiac Resynchronization Therapy, known as the WiSE-CRT device system as well as requiring a pacemaker implant which will also be leadless. Some of these participants may also require an AV Node ablation.
The purpose of this study is to assess the safety and efficacy of these two cardiac devices implanted in the order of operator preference (all implants undertaken on the same day or in a number of up to 4 separate sequenced implants/procedures).
详细描述
Heart Failure is a serious condition in which the heart is unable to pump enough blood to meet the body's demands. Some heart failure occurs when the heart chambers pump in a desynchronised manner, reducing the amount of blood pumped with each heartbeat.
The National Institute for Health and Care Excellence (NICE) recommends Cardiac Resynchronisation Therapy (CRT) for a subset of these patients. Patients undergoing CRT receive a special implantable pacemaker with an additional lead via the coronary sinus (CS) to synchronise the pumping of the heart chambers (ventricles). CRT works by pacing both the left & right ventricles to ensure they pump at the same time. This in turn, reduces heart failure symptoms.
CRT is an effective treatment for many patients, but limitations prevent some patients from benefiting:
- 5-10% of patients are considered untreated due to acute/chronic issues with the CS lead
- 30% of patients do not respond/improve
The WiSE-CRT System is an implantable pacing system capable of delivering pacing energy to the heart without using a pacing lead. It was designed to address the persistent limitations of current CRT systems. The WiSE-CRT System operates alongside a co-implanted pacemaker or CRT system and replaces the pacing function of the CS lead.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Male or Female, aged 18 years or above. Participant is willing and able to give informed consent for participation in the study.
- •Additionally, participants will be enrolled and classified from one of two distinct groups:
- •GROUP A: De novo totally leadless CRT
- •implant in whom the physician believes a totally leadless approach would be beneficial (e.g. wish to avoid lifelong transvenous lead implant, anatomical constraints, history of device infection). Examples of this could include:
- •symptomatic atrial fibrillation (AF) and an uncontrolled heart rate who are candidates for AV node ablation (irrespective of QRS duration and ejection fraction), in whom the physician believes a leadless pacing approach would be beneficial.
- •high degree atrioventricular block who has an indication for permanent pacing, with an ejection fraction ≤50%, are expected to require ventricular pacing more than 40% of the time and in whom the physician believes a leadless approach would be beneficial.
- •Other indications for leadless pacing including venous obstruction, pocket issues, infection risk (eg: chronic dialysis), etc.
- •GROUP B: Upgrade chronic intracardiac pacemaker to CRT
- •o Patient with existing intracardiac pacemaker with greater than 20% right ventricular (RV) pacing, who have developed symptomatic heart failure.
排除标准
- •Female participant who is pregnant, lactating or planning pregnancy during the course of the study
- •Inability to comply with the study follow-up or other study requirements
- •History of chronic alcohol/drug abuse and currently using alcohol/drugs
- •Non-ambulatory (or unstable) NYHA class 4
- •Life expectancy less than 12 months
- •Any other significant disease or disorder which, in the opinion of the Investigator, may either put the participants at risk because of participation in the study, or may influence the result of the study, or the participant's ability to participate in the study.
- •Enrolled in another clinical study that could confound the results of this study (note: patients enrolled in complementary study are eligible for enrolment)
结局指标
主要结局
Safety from major complications post device implants & procedures
时间窗: 6 months
Freedom from major complications following device implants and procedures, whether single stage or multiple staged procedures. Major complications are defined as any complication related to the device or procedure that results in death, hospitalization, prolongation of hospitalisation for 48 or more hours, permanent loss of device function due to a mechanical or electrical malfunction or system change (removal, repositioning, or replacement).
Efficacy of Cardiac Resynchronisation Therapy (CRT) measured by confirmation of Bi Ventricular Pacing on 12 lead ECG.
时间窗: 6 months
Evidence of biventricular pacing on 12 lead ECG recorded at 6 months will confirm CRT delivered by implanted devices for each participant.
次要结局
- Clinical Response of Each Participant measuring Left Ventricular End Systolic Volume.(6 months)
- Clinical response of Each Participant measuring New York Heart Association classification.(6 months)
- Clinical response of Each Participant measuring 6 Minute Walk Test(6 months)
- Clinical Response of Each Participant measuring Left Ventricular Ejection Fraction.(6 months)