简要总结

Head and neck cancer has an incidence of 600,000 cases per year worldwide (6th among cancers) and cause 380,000 death per year. Surgical resection is guided by the surgeon's inspection and palpation of the tumor. The prognosis of patients depends on the quality of the lesion's resection. The ENDOSWIR dispositive developed by engineers of Atomic Energy Comity of Grenoble allows for reliable and accurate analysis of tumor margins.

The camera provides a spectral analysis of the tissues according to their composition. We want to show that the prognosis of patients is improved by the use of this device. For that we need to prove first that it is not toxic for tissue analysis.

详细描述

ENDOSWIR dispositive is a InGaAs (Indium Gallium Arsenide) camera working at ambient temperature. The light is a quartz-tungstene-halogeno (QTH) source and the collected beam is filtered to illuminate in different precise wavelength bands in the short wave domain.

We will use it on tonsil tissue just after tonsillectomy and just before the histological analysis.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Tonsillectomy

排除标准

  • suspicion of Head-and neck cancer

结局指标

主要结局

Tissue integrity

时间窗: 1 day

Histological score : evaluation of cellular cohesion and denaturation of epithelial cells and conjunctive support tissue

次要结局

未报告次要终点