Clinical Investigation of the WaveLight® EX500 Excimer Laser for Hyperopic LASIK
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 44
- 试验地点
- 10
- 主要终点
- Percentage of Eyes With Manifest Refraction Spherical Equivalent (MRSE) Within ±0.50 Diopter (D) at Refractive Stability
研究概览
简要总结
The purpose of this study is to collect efficacy and safety data on the WaveLight EX500 excimer laser system for the correction of hyperopia with and without astigmatism by laser in situ keratomileusis (LASIK) treatment.
详细描述
Qualified subjects will receive LASIK treatment in both eyes and be followed for 1 year. Subjects will be asked to attend a total of 9 visits (Screening, Surgery, Day 1, Week 1, Month 1, Month 3, Month 6, Month 9, and Month 12). Total expected duration of subject participation is approximately 1 year.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Intended to treat bilaterally;
- •Hyperopia with or without astigmatism as specified in the protocol;
- •Stable vision as specified in the protocol;
- •Other protocol-defined inclusion criteria may apply.
排除标准
- •Glaucoma;
- •Cataracts;
- •Previous eye surgery;
- •Intent to have monovision treatment;
- •Other protocol-defined exclusion criteria may apply.
结局指标
主要结局
Percentage of Eyes With Manifest Refraction Spherical Equivalent (MRSE) Within ±0.50 Diopter (D) at Refractive Stability
时间窗: Up to Month 12 postoperative (Month 1 to Month 3, Month 3 to Month 6, Month 6 to Month 9, and Month 9 to Month 12)
A manifest refraction (manual refraction) was performed using a phoropter. The spherical equivalent was calculated by adding the sum of the sphere power with half of the cylinder power. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at 3 months after surgery when compared with the 1-month visit) that meets all the protocol-specified stability requirements. For this endpoint, change in MRSE by less than or equal to 0.50 D was reported. Both eyes contributed data to this analysis.
Percentage of Eyes With MRSE Within ±1.00 D at Refractive Stability
时间窗: Up to Month 12 postoperative (Month 1 to Month 3, Month 3 to Month 6, Month 6 to Month 9, and Month 9 to Month 12)
A manifest refraction (manual refraction) was performed using a phoropter. The spherical equivalent was calculated by adding the sum of the sphere power with half of the cylinder power. Refractive stability was defined as the latter of two postoperative manifest refractions performed at least 3 months apart (or at 3 months after surgery when compared with the 1-month visit) that meets all the protocol-specified stability requirements. For this endpoint, change in MRSE by less than or equal to 1.00 D was reported. Both eyes contributed data to this analysis.
Percentage of Eyes Experiencing Ocular Adverse Events (AEs)
时间窗: Up to Month 12 postoperative
An AE was defined as any untoward medical occurrence, unintended disease or injury, or untoward clinical signs (including abnormal laboratory findings) in subjects, users, or other persons, whether or not related to the investigational medical device (test product). Ocular adverse events were collected for this outcome measure. No hypothesis testing was prespecified for this endpoint.
次要结局
未报告次要终点