In Vivo Testing for SpO2 Accuracy on Healthy Adults Under Laboratory Conditions
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 12
- 试验地点
- 2
- 主要终点
- Measure of Accuracy (ARMS [%])
研究概览
简要总结
The goal of this clinical trial is to compare values of SpO2 provided by the DUT and the SaO2 used as a reference in healthy adults. The main question it aims to answer is: what is the accuracy of the subject device compared to the 510(k)-cleared reference device
A radial arterial cannula was placed in the wrist of each subject for arterial blood sampling and a subject device was placed on each arm of the patients. Then, hypoxia was induced to different stable levels of oxyhemoglobin saturation between 70-100% by having subjects breathe mixtures of nitrogen, air, and carbon dioxide controlled by the study physician. Values of SpO2 and SaO2 were then compared to provide a root mean square error and conclude on the subject device accuracy.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Other
- 盲法
- None
入排标准
- 年龄范围
- 21 Years 至 50 Years(Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Healthy adult
- •Normal hemoglobin levels (Hemoglobin > 10 g/dL)
排除标准
- •Systemic illness
结局指标
主要结局
Measure of Accuracy (ARMS [%])
时间窗: 1 hour
Arms, given in percentage or rms error, is the calculation of root mean square between SpO2 values from the subject device and the SaO2 from the Hemoximeter. It is a way of averaging the absolute values of errors over the full measurement range and is a similar measurmeent to a "rms error". It indicates how far (accurate) the provided value is from the reference used.
次要结局
未报告次要终点