In Vivo Testing for SpO2 Accuracy on Healthy Adults Under Laboratory Conditions

Gabi SmartCare2 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2023年2月28日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
12
试验地点
2
主要终点
Measure of Accuracy (ARMS [%])

研究概览

简要总结

The goal of this clinical trial is to compare values of SpO2 provided by the DUT and the SaO2 used as a reference in healthy adults. The main question it aims to answer is: what is the accuracy of the subject device compared to the 510(k)-cleared reference device

A radial arterial cannula was placed in the wrist of each subject for arterial blood sampling and a subject device was placed on each arm of the patients. Then, hypoxia was induced to different stable levels of oxyhemoglobin saturation between 70-100% by having subjects breathe mixtures of nitrogen, air, and carbon dioxide controlled by the study physician. Values of SpO2 and SaO2 were then compared to provide a root mean square error and conclude on the subject device accuracy.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
21 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Healthy adult
  • Normal hemoglobin levels (Hemoglobin > 10 g/dL)

排除标准

  • Systemic illness

结局指标

主要结局

Measure of Accuracy (ARMS [%])

时间窗: 1 hour

Arms, given in percentage or rms error, is the calculation of root mean square between SpO2 values from the subject device and the SaO2 from the Hemoximeter. It is a way of averaging the absolute values of errors over the full measurement range and is a similar measurmeent to a "rms error". It indicates how far (accurate) the provided value is from the reference used.

次要结局

未报告次要终点