临床试验/NCT03085251
NCT03085251已完成不适用

Clinical System Accuracy Evaluation With OneTouch® Verio Flex Blood Glucose Monitoring System (BGMS)

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm2 个研究点 分布在 1 个国家目标入组 109 人开始时间: 2017年3月16日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
109
试验地点
2
主要终点
System accuracy criteria

研究概览

简要总结

The objective of this study is the evaluation of system accuracy based on EN ISO 15197:2015 (ISO 15197:2013), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. ISO 15197:20132 was harmonized with the regulations of the European Union as EN ISO 15197:2015. This harmonization had no impact on the requirements and procedu¬res in ISO 15197:2013; changes were made to the foreword and an informative annex. In this study, system accuracy evaluation will be performed for OneTouch® Verio Flex (LifeScan Europe) on behalf of LifeScan Scotland Limited (customer) with three different reagent system lots.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Male or female, with type 1 diabetes, type 2 diabetes (or subjects without diabetes)
  • Signed informed consent form
  • Minimum age of 18 years
  • Subjects are legally competent and capable to understand character, meaning and consequences of the study.
  • If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
  • Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
  • Signature of subjects to document consent with these procedures on informed consent form.

排除标准

  • Pregnancy or lactation period
  • Severe acute disease (at the study physician's discretion)
  • Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
  • Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
  • Being unable to give informed consent
  • < 18 years
  • Legally incompetent
  • Being committed to an institution (e.g. psychiatric clinic)
  • Language barriers potentially compromising an adequate compliance with study procedures
  • Dependent on investigator or sponsor
  • If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy (concentration category 2), subjects with type 1 diabetes, suffering from:
  • Coronary heart disease
  • Condition after myocardial infarction
  • Condition after cerebral events
  • Peripheral arterial occlusive disease
  • Hypoglycemia unawareness

结局指标

主要结局

System accuracy criteria

时间窗: For each subject, the experimental phase has an expected duration of up to 6 hours

Based on EN ISO 15197:2015 (ISO 15197:2013) (E), the BGMS shall meet both of the following criteria: Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations \< 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l). Each lot shall pass acceptability criterion A. Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.

次要结局

未报告次要终点

研究者

发起方
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
申办方类型
Other
责任方
Sponsor