临床试验/NCT02531386
NCT02531386已完成不适用

System Accuracy Evaluation of Accu-Chek Aviva Nano (Roche Diagnostics), Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan), Contour Next Link 2.4 (Bayer Consumer Care), One Additional CE-marked Blood Glucose Monitoring System Following ISO 15197:2013

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm2 个研究点 分布在 1 个国家目标入组 120 人开始时间: 2015年8月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
120
试验地点
2
主要终点
System accuracy criteria (see description)

研究概览

简要总结

The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. In this study, system accuracy evaluation will be performed with CE-marked BGMS (BGMS = Blood Glucose Monitoring System) common among insulin pump users - Accu-Chek Aviva Nano and Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan) Contour Next Link 2.4 (Bayer Healthcare Diabetes Care) and Accu-Chek Mobile with three different reagent system lots for each BGMS.

The study will be performed in 2 parts. Each part will comprise testing of 3 BGMS.

研究设计

研究类型
Interventional

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Male or female, with type 1 diabetes, type 2 diabetes or no diabetes
  • For provoked blood glucose excursions due to insulin dose adjustment:
  • Male or female with type 1 diabetes or
  • Type 2 diabetes and intensified insulin therapy or insulin pump therapy.
  • Minimum age of 18 years
  • Signed informed consent form
  • Legally competent and capable to understand character, meaning and consequences of the study

排除标准

  • Pregnancy or lactation period
  • Severe acute disease (at the study physician's discretion)
  • Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
  • Physical or mental constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
  • Incapability of giving informed consent
  • < 18 years
  • Legally incompetent
  • Accommodated in an institution (e.g. psychiatric clinic)
  • Language barriers potentially compromising an adequate compliance with study procedures
  • Dependent from investigator or sponsor
  • For provoked blood glucose excursions 50 - 80 mg/dl (concentration category 2):
  • Coronary heart disease
  • Condition after myocardial infarction
  • Cerebral incidence
  • Peripheral arterial occlusive disease
  • Hypoglycemia unawareness

结局指标

主要结局

System accuracy criteria (see description)

时间窗: For each subject, the experimental phase has an expected duration of up to 6 hours

System accuracy criteria Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria: Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations \< 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l). Each lot shall pass acceptability criterion A. Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes. Applying ISO 15197:2003 accuracy criteria: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations \< 75 mg/dl (4.2 mmol/l) or within ± 20 % at glucose concentrations ≥ 75 mg/dl (4.2 mmol/l).

次要结局

未报告次要终点

研究者

发起方
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
申办方类型
Other
责任方
Sponsor