Trauma Equivalency Study of the CORA® and TEG® 5000 Systems
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 325
- 试验地点
- 3
- 主要终点
- Composite measure: Blood coagulation and hemostasis parameters R (min), C-ACT (min), K (min), Angle (degrees), MA (mm) and LY30 (%) parameters for both TEG 5000 and CORA systems
研究概览
简要总结
This study is being performed to obtain data about the CORA System when used in a trauma clinical setting. The CORA analyzer is a new device that is currently being tested in trauma clinical applications and is not yet cleared for this purpose by the Food and Drug Administration (FDA) for sale in the United States.
详细描述
The purpose of the study is to validate the CORA System by comparing the results of this test to that of the standard of care device called the TEG 5000 system. Both systems evaluate hemostasis, a highly complex and interactive process involving many protein and cellular elements, especially platelets, occurring in a series of interrelated phases, which results in either normal hemostasis or a tendency toward bleeding or clotting complications.
The study protocol specifies testing using multiple reagents which are similar for both the CORA and TG 5000 systems. Tests are run in duplicate on blood samples taken from patients undergoing treatment for injuries and trauma. Following the protocol defined in CLSI Standard EP09-A2, equivalence between the two systems will be demonstrated if the conclusions from the Standard Section 7, Interpreting Results and Comparing to Internal Performance Criteria, are found to be acceptable.
All blood samples drawn for this study will use venous blood. Only citrated or heparinized whole blood samples are used. All samples are tested twice (run in duplicate). Quality Control procedures are run on both instruments at a frequency specified by the manufacturers. Results obtained from this testing are not used to make treatment decisions.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Cross Sectional
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •18 years of age and older
- •Complete patient history including age, race, gender and type of trauma should be documented
排除标准
- •Patients under the age of 18
- •Native whole blood will not be used in this study
结局指标
主要结局
Composite measure: Blood coagulation and hemostasis parameters R (min), C-ACT (min), K (min), Angle (degrees), MA (mm) and LY30 (%) parameters for both TEG 5000 and CORA systems
时间窗: 1 day, measured at time of trauma treatment; no follow-up
次要结局
未报告次要终点