简要总结
The purpose of this study is to compare difference of displacement between Polyvinyl Chloride(PVC) and Silicon double-lumen endobronchial tubes(DLTs) during changing position from supine to lateral in thoracic surgery.
详细描述
Lung isolation and differential lung ventilation, resulting in collapse of operative lung and ventilation of non-operating lung (one lung ventilation; OLV) are essential for thoracic surgery in lung, esophagus and thoracic aorta.
DLT, consisted of tracheal and bronchial lumens, is commonly used to perform differential lung ventilation. The correct position of each lumen in trachea and main bronchus is essential for successful OLV. However, DLT displacement occurs commonly while patients are changed position from the supine to the lateral. The deleterious consequences of a malpositioned DLT can be substantial, even life-threatening including severe hypoxemia.
Considering displacement of DLT during position change and surgical traction, previous studies recommended endobronchial cuff of DLT should be positioned at 0.5cm below the tracheal carina when using left-sided DLT. However, it is based on the data from several research conducted by PVC DLT.
These days, Silicon DLT, different in material from PVC DLT, is also widely used in clinical practice. Silicon DLT is different in not only position of ballon but also distance of each lumen. Furthermore,there are no identified studies about the displacement of silicon DLT during patient position change in thoracic surgery.
Therefore, it is necessary to measure the movement of silicon DLT during positional change from supine to lateral for finding the proper positioning of DLT.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Prevention
- 盲法
- Single (Participant)
入排标准
- 年龄范围
- 18 Years 至 75 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patient required differential lung ventilation in elective thoracic surgery American society of Anesthesiologist physical status(ASA PS) 1~3
排除标准
- •emergency surgery, difficult intubation, poor lung function to accomplish OLV during surgery
结局指标
主要结局
the Incidence of Clinically Significant Displacement of DLTs During Change of Patient Position
时间窗: from supine to lateral decubitus position during surgery
The tracheal distance was defined as the distance between the distal tip of the tracheal lumen and tracheal carina whereas the bronchial distance was defined as that between the bronchial carina and distal tip of endobronchial lumen. The displacement of DLT was determined by changes in tracheal and bronchial distances, obtained by subtracting supine measurements from lateral measurements and subtracting measurements at start of surgery from measurements at the end of surgery, respectively. Clinically significant displacement was defined when the DLT was deviated by more than 10 mm from the initial correct position, regardless of the direction of displacement.
次要结局
- the Incidence of Critical Malposition of DLTs(from supine to lateral decubitus position)