Comparison of UESCOPE VL 400, I-VIEW, Airtraq Videolaryngoscopes and Macintosh Laryngoscope for Tracheal Intubation in Simulated Out-of-Hospital Conditions: A Randomized Crossover Manikin Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 30
- 试验地点
- 1
- 主要终点
- time to ventilation
研究概览
简要总结
In the study, three videolaryngoscopes: UESCOPE VL 400 (Zhejiang UE Medical Corp. Zhejiang, China), I-View™ VL (Intersurgical Ltd, Wokingham, Berkshire, UK), Airtraq sp® (Prodol Meditec S.A, Vizcaya, Spain) were compared with the most widely used, classic Macintosh laryngoscope (HEINE Optotechnik GmbH & Co. KG, Gilch-ing, Germany) under simulated out-of-hospital conditions by people without clinical experience on a manikin model (Laerdal Airway Management Trainer Stavanger Norway manikin of universal difficulty), while using an endotracheal tube without and with a stylet. Simulated out-of-hospital conditions were created by placing the manikin at floor level. A random selection of 30 students of the third year of Emergency Medicine, full-time first-degree program at the Medical University of Lodz, Poland, were enrolled in the study. The following were assessed: intubation efficiency, time to ventilation (primary endpoints); intubation effort, intubator comfort, tooth damage and visibility of the glottis according to the Cormack-Lehane scale (secondary endpoints).
详细描述
Pre-study preparation All participants attended a 30-minute lecture on laryngoscope construction, principles of usage, anatomical structure of the airways, and intubation techniques. Following the lecture, the instructor demonstrated proper intubation using each of the four laryngoscopes that were being tested. Next, the students participated in a supervised workshop where they had the opportunity to intubate a mannequin placed on an operating table at an optimal height using each of the tested laryngoscopes with and without a stylet. One month later, 30 students participated in the actual study.
Simulation model To create simulated out-of-hospital conditions, a certified airway training manikin (Laerdal Airway Management Trainer Stavanger Norway of universal difficulty) was placed in a neutral position at floor level.
The trial tested three unchanneled videolaryngoscopes: UESCOPE VL 400 (Zhejiang UE Medical Corp. Zhejiang, China), I-View™ VL (Intersurgical Ltd, Wokingham, Berkshire, UK), Airtraq sp ® (Prodol Meditec S.A, Vizcaya, Spain), and a Macintosh laryngoscope (HEINE Optotechnik GmbH & Co. KG, Gilch-ing, Germany) (see Fig. 1).
Each participant performed eight intubations on the manikin using all four devices. The intubation order was randomized using a blocked randomization strategy generated by the Randomizer Program (randomizer.org). A sealed opaque envelope was used to determine the order in which the laryngoscopes were used, and whether the tube was with or without a stylet.
After each intubation, there was a minimum 10-minute pause before the next one. The subject then proceeded to the next intubation with the randomly selected laryngoscope, with the tube and stylet as determined by the sealed envelope.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Crossover
- 时间视角
- Cross Sectional
入排标准
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •attaneding students
排除标准
- •lack of conscent to participate
结局指标
主要结局
time to ventilation
时间窗: during the procedure
time from grabbing the laryngoscope to confirmed intubation measured in sec
intubation efficacy
时间窗: during the procedure
endotracheal intubation confirmed by ventilation of manikin's lungs, measured yes/no
次要结局
- visibility of the glottis according to the Cormack-Lehane scale(during the procedure)
- intubator comfort,(during the procedure)
- intubation effort(during the procedure)
- tooth damage(during the procedure)
研究者
Tomasz Gaszynski
prof. dr hab.
Medical University of Lodz