跳至主要内容
临床试验/NCT03496493
NCT03496493已完成不适用

Perioperative Comparison of the Agreement Between a Portable Fingertip Pulse Oximeter Versus Conventional Bedside Pulse Oximeter in Adult Patients

University of Cape Town1 个研究点 分布在 1 个国家目标入组 220 人开始时间: 2018年4月15日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
220
试验地点
1
主要终点
Saturation measurement agreement

研究概览

简要总结

Perioperative non-invasive measurement of a patient's peripheral arterial oxygen saturation has become an accepted standard of care endorsed by anaesthesiologists and their regulatory bodies throughout the world. Pulse oximeters are an integral item on the World Health Organisation's Surgical Safety Checklist which is performed prior to the commencement of every surgical procedure. They are also utilised by other medical professionals and patients in various scenarios. When compared to conventional bedside pulse oximeters, portable fingertip devices have the advantages of cost-effectiveness, high portability, ease of use and battery operation. A review of recent literature reveals a paucity of accuracy studies in adult patients with various comorbidities in the clinical setting. Most data has been obtained under ideal laboratory conditions utilizing healthy adult volunteers. This study aims to pragmatically investigate the performance of a portable fingertip pulse oximeter in adult patients in a hospital setting.

详细描述

The purpose of this prospective, quantitative study is to compare the agreement between the measured arterial oxygen saturation levels using a portable fingertip versus a conventional bedside pulse oximeter in adult surgical patients, presenting for elective and emergency surgery to Groote Schuur Hospital over a four-week period. Adult surgical patients, not meeting any exclusion criteria, who present for elective or emergency surgery will be recruited. 200 data pairs (SpO2, heart rate and skin tone) will be recorded in the perioperative period (operating rooms, recovery room and intensive care unit).

Existing monitoring devices will not be disturbed. The agreement between the two devices will be compared using a Bland-Altman plot, Microsoft Excel and SPSS software. A root mean square difference value will also be calculated from these values.

No sensitive or personal patient information will be recorded, and pulse oximetry is essential harmless. Potential cross-infection will be avoided by excluding patients with known communicable diseases or open wounds/ bodily fluids in the area to have the oximeter probes applied. The study devices will be disinfected after each use. Recruitment will not affect clinical care.

The significance of this study into the performance of portable fingertip pulse oximeters in adult surgical patients will result in the availability of cheaper devices, deemed as vital anaesthetic equipment by the WHO and various anaesthetic regulatory bodies, in lower income medical facilities. It will also give anaesthetists and other medical personal the confidence to make clinical decisions based on these highly portable devices, thereby making pulse oximeters more available in resource limited areas. It will also add further data to existing studies and help assess whether darker skin pigment affects the performance of these devices in the clinical setting, which is highly relevant in an African country like South Africa.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adult surgical patients presenting for elective or emergency surgery

排除标准

  • Contact precautions due to high risk of transmissable infectious disease
  • Significant hypotension or hypoperfusion (systolic blood pressure <80 mmHg, or hypothermia)
  • Motion artefacts due to excessive patient movements
  • Known presence of variant haemoglobin species (eg. carboxy- or methaemoglobin)
  • Intraoperative dye use (eg. methylene blue or indocyanine green)
  • Nail polish or black henna
  • Tape or bandages over the fingers
  • Absent/inadequate pulse oximetry tracing on existing monitor

结局指标

主要结局

Saturation measurement agreement

时间窗: 1 minute

Bland-Altman analysis of saturation measurement pairs, with a priori acceptable limits of agreement (1.96 x SD measurement differences) within 3%. This is in accordance with ISO and FDA guidelines for testing of these devices (see relevant standard).

次要结局

  • Average root mean square difference(1 minute)
  • Skin tone effect(1 minute)
  • Perfusion (pulse magnitude) effect on oximeter agreement(1 minute)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Ross Hofmeyr

Associate Professor, Dept Anaesthesia & Perioperative Medicine

University of Cape Town

研究点 (1)

Loading locations...

相似试验