The PainFree SST Clinical Study: Clinical Study to Evaluate System Safety and Clinical Performance of the Protecta ICD + CRT-D and to Assess the Capacity of the Protecta Features to Reduce Inappropriate Shocks.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 2,770
- 试验地点
- 122
- 主要终点
- Percentage of Subjects Who Are Inappropriate Shock Free
研究概览
简要总结
The purpose of the study is two-fold. In Phase I (Protecta Clinical Study), system performance will be evaluated. In Phase II (PainFree SST), the inappropriate shock-free rate at one year of subjects implanted with a Medtronic Protecta implantable cardioverter defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) will be evaluated.
详细描述
The PainFree SST study was a prospective, multicenter clinical trial, which was conducted in two consecutive phases. Phase I assessed safety as determined by assessment of delays in ventricular fibrillation (VF) arrhythmia detection time using new algorithms. Phase II assessed the influence of new algorithms on inappropriate and unnecessary shocks at 1 year of follow-up. Subjects enrolled in Phase I continued in Phase II of the study and data from all enrolled subjects contributed to the analysis of Phase II objectives. A sub-study of Phase II randomized secondary prevention subjects to either 18/24 NID or 30/40 NID in order to assess the safety of prolonged detection in these patients.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients meeting one of the following criteria can be included in Phase I of the study:
- •Patients indicated (per local indications) for a dual chamber implantable cardioverter defibrillator (DR-ICD)/cardiac resynchronization therapy defibrillator (CRT-D)
- •Patients undergoing a device replacement or upgrade to DR-ICD or CRT-D and with the protocol required leads (patients must be enrolled prior to implant)
- •Patients meeting one of the following criteria can be included in Phase II of the study:
- •Patients indicated for a DR-ICD/VR-ICD/CRT-D device and intended to receive a Protecta device
- •Patients undergoing a DR-ICD/VR-ICD/CRT-D device replacement or an upgrade to a Protecta DR-ICD/VR-ICD or CRT-D
- •Patients indicated for a DR-ICD/VR-ICD/CRT-D already implanted with a Protecta device and the protocol required leads, who have not yet been discharged after implant
排除标准
- •Patients with a mechanical tricuspid heart valve
- •Patients enrolled or intended to participate in any concurrent drug and/or device study, which would confound the results of this trial as determined by Medtronic, during the course of this clinical study
- •Patients with medical conditions that preclude the testing required for all patients by the study protocol or that otherwise limit study participation required for all patients
- •Patients with exclusion criteria required by local law (e.g. age, pregnancy, breast feeding, etc)
- •Patients anticipated not being able to complete the study
- •Patients unwilling to provide written informed consent
研究组 & 干预措施
Single-chamber detetction
Patients implanted with a Protecta VR-ICD.
干预措施: Protecta VR-ICD (Device)
Dual-chamber detection
Patients implanted with a Protecta DR-ICD or CRT-D.
干预措施: Protecta DR-ICD or CRT-D (Device)
结局指标
主要结局
Percentage of Subjects Who Are Inappropriate Shock Free
时间窗: Implant to one year post-implant
Primary objective of Phase II. Subjects implanted with a VR device will be analyzed separately from subjects implanted with a DR / CRT-D device. An inappropriate shock is a shock delivered by the defibrillator when the patient's heart rhythm was not a tachyarrhythmia, as adjudicated by the independent Episode Review Committee .
Percentage of Subjects With Unanticipated Severe Adverse Device Effects (Phase I)
时间窗: Implant to one month post-implant
In Phase I, only DR-ICD/CRT-D devices were implanted, so that for Phase I objectives there is only 1 arm to report results for.
Percentage of Phase I Subjects Where the Ventricular Fibrillation (VF) Detection Time With Protecta Features on is no More Than 2 Seconds Longer Than the VF Detection Time With Protecta Features Off
时间窗: At implant
In Phase I, only DR-ICD/CRT-D devices were implanted, so that for Phase I objectives there is only 1 arm to report results for.
次要结局
- Percentage of Secondary Prevention Subjects Who Are Syncopal Event Free(Implant to one year post-implant)