临床试验/NCT05863091
NCT05863091招募中不适用

Respiratory Rate Validation Study With Accuracy Analysis of Respiratory and Breathing Pattern Extracted by Impedance Change of Patch-type Electrocardiogram Device (HiCardi+ Wearable Patch Device, Mezoo Co., Ltd.)

Yonsei University1 个研究点 分布在 1 个国家目标入组 211 人开始时间: 2023年3月9日最近更新:
适应症

试验速览

阶段
不适用
状态
招募中
入组人数
211
试验地点
1
主要终点
Respiratory rate measured by pulmonary function test

研究概览

简要总结

The purpose of this study is to analyze the accuracy of respiratory and breathing patterns generated through impedance changes generated by a patch-type electrocardiogram device (HiCardi+ wearable patch device, Mezoo Co., Ltd.), targeting patients undergoing pulmonary function testing and ventilator application.

详细描述

Subjects The target number of participants for this study is 211, consisting of 181 patients undergoing pulmonary function testing and 30 patients using a ventilator.

  • Spontaneous breathing subjects (181 patients): Assuming that a pulmonary function testing examiner with spontaneous breathing would examine about 50 patients per day, a total of 181 subjects will be targeted, taking into consideration the possibility of refusals.
  • Non-Spontaneous breathing subjects (30 patients): In addition, in the case of patients with a Richmond Agitation Sedation Scale of -4 to -5 who use a ventilator and do not have spontaneous breathing, the final 30 subjects will be conducted for the purpose of data collection for about 6 hours.

Focus of the study The tidal volume and respiratory amplitude will be analyzed in detail through pulmonary function tests, and the respiratory rate will be analyzed through ventilator data.

Study method The study method is as follows: In the pulmonary function testing room, consent will be obtained from patients undergoing pulmonary function testing, and a patch-type electrocardiogram device will be attached before the test is conducted. The records of the ECG patch generated during the approximately 30-minute pulmonary function test and the results of the pulmonary function test will be acquired. Based on the impedance obtained from the patch-type electrocardiogram, the pulmonary function test results are extracted as the reference data related to the respiratory pattern. Correlation analysis is performed on the tidal volume and respiratory amplitude obtained through this comparison. Additionally, Bland-Altman analysis is conducted on the validation set, which is divided into an 8:2 training set and validation set, for the algorithm that predicts tidal volume and respiratory amplitude.

For patients with RASS -4 to -5 who are on a ventilator, the patch-type electrocardiogram device is attached for about 6 hours after the consent of the patient caregiver, and the data generated from the ECG patch and the ventilator are compared and analyzed. Based on the impedance obtained from the patch-type electrocardiogram, the data from the ventilator will be used as reference data related to respiratory rate. Mean Difference (MD), Standard Deviation (SD), degrees of freedom (df), bias, regression, and limits of agreement will be calculated through correlation analysis and Bland-Altman analysis on the obtained respiratory rate.

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Parallel
主要目的
Other
盲法
None

入排标准

年龄范围
19 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adults 19 and older

排除标准

  • patients under the age of 19
  • patients cannot apply the patch-type electrocardiogram device due to physical defects or contact allergic reactions
  • patients do not consent to participate in the study

结局指标

主要结局

Respiratory rate measured by pulmonary function test

时间窗: During the pulmonary function test (about 30 minutes)

The respiratory rate will be analyzed for correlation and Bland-Altman analysis using continuous variables generated from pulmonary function testing, ventilator information, and patch-type ECG device. Mean difference (MD), standard deviation (SD), degrees of freedom (df), bias, regression, and limits of agreement will be calculated.

Respiratory rate measured by ventilator

时间窗: about 6 hours

The respiratory rate will be analyzed for correlation and Bland-Altman analysis using continuous variables generated from pulmonary function testing, ventilator information, and patch-type ECG device. Mean difference (MD), standard deviation (SD), degrees of freedom (df), bias, regression, and limits of agreement will be calculated.

Respiratory rate measured by patch-type electrocardiogram device

时间窗: up to 6 hours

The respiratory rate will be analyzed for correlation and Bland-Altman analysis using continuous variables generated from pulmonary function testing, ventilator information, and patch-type ECG device. Mean difference (MD), standard deviation (SD), degrees of freedom (df), bias, regression, and limits of agreement will be calculated.

次要结局

  • Tidal volume measured by pulmonary function test, ventilator, and patch-type electrocardiogram device(-Pulmonary function test: During the pulmonary function test (about 30 minutes) -Ventilator application: about 6 hours)
  • respiratory amplitude measured by pulmonary function test, ventilator, and patch-type electrocardiogram device(-Pulmonary function test: During the pulmonary function test (takes about 30 minutes) -Ventilator application: about 6 hours)