Evaluation of Nucleus 6 Sound Coding Algorithms
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- Cochlear
- 入组人数
- 97
- 主要终点
- Percentage of Words Correct on AzBio Sentences in Noise S0N90
研究概览
简要总结
The purpose of the study was to evaluate the features of the Nucleus 6 Sound Processor (CP900 series) on existing Nucleus recipients who currently use the previous generation sound processor, the Cochlear Nucleus CP810 (also commonly known as the Nucleus 5), with a Cochlear Nucleus CI24RE, CI512 or CI422 cochlear implant. The features evaluated included an enhanced Automatic Gain Control (eAGC), the option to provide acoustic amplification in combination with traditional electric stimulation through an acoustic component, and three input processing algorithms: Signal to Noise Ratio - Noise Canceller (SNR-NC), Wind Noise Reduction (WNR) and Automatic Scene Analyzer (SCAN).
As is customary for a single-subject research design, study results will be analyzed for each subject individually. For evaluations where testing took place under two conditions (i.e. Endpoint 2, 3, 4, and 5) a non-inferiority hypothesis test based on the paired comparisons of the % of words correct will be performed. For each subject, the paired difference (P1 minus P2) will be calculated and the mean difference will be tested statistically against a non-inferiority null hypothesis of 10%. Successful rejection of the null hypothesis allows for a conclusion that the performance under the test condition 2 (based on the Nucleus 6 sound processor) will be non-inferior to the test condition 1 (based on the Nucleus 5 sound processor).
Ho: P1 - P2 ≥ 10% Ha: P1 - P2 < 10%
We hypothesize acceptance of the null hypothesis for the Nucleus 6 features.
Testing will be based on a paired t-test with a one-sided 0.05 alpha level. The value of 10% was selected as the non-inferiority margin as one that can be tested with reasonable power and was clinically meaningful.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 8 Years 至 —(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Eight years of age or older
- •CI24RE, CI500 or CI422 series Nucleus cochlear implant types (commercially approved)
- •At least 3 months experience with the cochlear implant
- •Post-linguistic onset of bilateral sensorineural hearing loss, with no congenital component to the hearing loss
- •Native speaker of English, the language used to assess speech perception performance
- •Willingness to participate in and to comply with all requirements of the protocol
排除标准
- •Unrealistic expectations on the part of the subject, regarding the possible benefits, risks, and limitations that are inherent to the prosthetic device
- •Unwillingness or inability of the candidate to comply with all investigational requirements
- •Additional handicaps that would prevent or restrict participation in the audiological evaluations
结局指标
主要结局
Percentage of Words Correct on AzBio Sentences in Noise S0N90
时间窗: Each assessment was completed acutely which equated to approximately 2 hours of testing
Individuals were tested with AzBio Sentences in Noise using a speech front (0 degrees) and noise presented to the implant ear (90 degrees) using the following features: enhanced Automatic Gain Control (eAGC), Signal to Noise Ratio - Noise Canceller (SNR-NC), Wind Noise Reduction (WNR) and Automatic Scene Analyzer (SCAN).
次要结局
未报告次要终点