Analytic Validation of the Optina Retinal Deep Phenotyping.TM (RDP) Platform

Optina Diagnostics Inc.1 个研究点 分布在 1 个国家目标入组 43 人开始时间: 2023年9月19日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
43
试验地点
1
主要终点
The reliability of the investigational device's imaging process will be analyzed using agreement of CAS Test results within 95% confidence intervals for the following conditions:

研究概览

简要总结

This is an observational study designed to assess the reliability and consistency of the Optina awAIrTM CAS Test for detecting phenotypic changes in the retina that are associated with beta-amyloid status (positive or negative). The purpose of this study is to ensure that the retinal image capture process is consistent and under control, regardless of the operator or the time of measurement. The study will include enough repeated measurements on different systems, by three operators, and at different times to demonstrate.

详细描述

The purpose of this study is to assess the reliability of the Optina Retinal Deep Phenotyping.TM(RDP) platform. The RDP Platform is comprised of the Optina-4CTM retinal camera and its Optina-4C ConnectTM software, as well as the Optina awAIrTM-CAS Test (also called CAS Test) for detecting phenotypic changes in the retina that are associated with cerebral beta-amyloid status (positive or negative). This study will quantify points of variability in the data input for the CAS Test, including:

  • Left eye vs. right eye (retinal scan)
  • Eye algorithm selection (eye assessment obtained from different eye specialists either an Ophthalmologist or an Optometrist)
  • Hardware variability (Inter-instrument Variability retinal scans obtained from different Optina-4CTM cameras)
  • Inter-operator variability (retinal scans obtained from different Users of the Optina-4CTM camera)
  • Intra-operator variability (retinal scans obtained by the same user on the same Optina-4CTM camera

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

年龄范围
50 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Male and female adults aged 50 years and older with vision in both eyes.
  • Sufficient fluency in English or French to participate in study visits.
  • Ability to give informed consent.
  • Individuals with both eyes meet the eligibility criteria for retinal imaging.

排除标准

  • Pupil dilation contraindicated (due to pathology, or presence of 3 quadrants with Van Herick grading of 0 or 1 without iridotomy).
  • Inadequate pupil dilatation (< 6mm diameter) preventing uniform illumination of the retina with the Optina 4CTM camera.
  • Diagnosis of glaucoma or signs of glaucoma (excavation ratio ≥0.7).
  • Signs of vascular occlusion or retinopathy (microaneurysm, exudate, hemorrhage, or edema) within a diameter of 10 mm from the mid-point between the optic nerve head and the macula.
  • Presence of drusen and/or age-related macular degeneration (AREDS 9-step scale
  • ≥4 - cumulative drusen area diameter ≥ 250 µm, pigmentary changes and cumulative drusen area diameter ≥ 63 µm or pigmentary changes and cumulative geographic atrophy area diameter ≥ 354 µm).
  • Nuclear sclerosis > 2 (LOCS II four-point grading system) or presence of central cortical or central posterior subcapsular cataract.
  • Deficient visual fixation (inability to fixate for at least 2 s).
  • Refractive error outside the range of -15 D to +15 D.
  • Corneal or media opacities (e.g., Weiss ring) affecting retinal imaging on a cumulative area > 1 disc area within a diameter of 10 mm from the mid-point between the optic nerve head and the macula (i.e., the area of interest for the Optina 4CTM imaging)
  • Scar, atrophy, naevus, tumor, epiretinal membrane or retinal pucker with a cumulative area > 1 disc area within a diameter of 10 mm from the mid-point between the optic nerve head and the macula.
  • Papilledema and/or optic disc swelling.
  • Macular anomaly (e.g., macular hole, dystrophy, degeneration).
  • Inability to obtain up to 5 images or a minimum of 3 images of satisfactory quality with the Optina 4CTM per the Optina Diagnostics quality index software and /or per the eye specialists' evaluation. This exclusion criterion also applies if participants are unable to provide the required number of images during the second visit as well.
  • Currently taking or has previously taken cerebral amyloid modifying medication.

结局指标

主要结局

The reliability of the investigational device's imaging process will be analyzed using agreement of CAS Test results within 95% confidence intervals for the following conditions:

时间窗: 12 months

* Eye to eye variability * Eye selection process variability * Instrument to instrument variability * Inter-Operator variability * Intra-Operator variability

次要结局

未报告次要终点

研究者

发起方
Optina Diagnostics Inc.
申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...

相似试验