M- HEALTHCARE Confirmatory Clinical Trials Used for Measurement of the ABP for the Accuracy and Safety Assessment of the Module Measuring Blood Pressure

Seoul National University Hospital2 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2016年12月最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
100
试验地点
2
主要终点
mean difference of systolic blood pressure

研究概览

简要总结

The accuracy and safety of blood pressure module of new mobile device (M- HEALTHCARE device) is evaluated according to Korean FDA guidelines in 100 healthy volunteers.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
20 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Participants whose systolic blood pressure <=110 should be at least 10% of whole cohort
  • Participants whose systolic blood pressure >=160 should be at least 10% of whole cohort
  • Participants whose diastolic blood pressure >=100 should be at least 10% of whole cohort
  • Participants whose diastolic blood pressure <=70 should be at least 10% of whole cohort
  • Participants whose arm circumference are between 20cm and 30cm should be 25 - 50% of whole cohort
  • Participants whose arm circumference are between 30cm and 40cm should be 50 - 75% of whole cohort

排除标准

  • heart disease with abnormal rhythm
  • peripheral vascular anomaly or disease
  • inadequate to be applied blood pressure cuff (e.g. trauma on arm)
  • pregnant or lactating women

研究组 & 干预措施

EXPERIMENTAL

Experimental

Blood pressure check using mobile device and control device

干预措施: blood pressure check (Device)

结局指标

主要结局

mean difference of systolic blood pressure

时间窗: 1 day at visit 2

mean difference of blood pressure between new and control device

mean difference of diastolic blood pressure

时间窗: 1 day at visit 2

mean difference of diastolic blood pressure between new and control device

standard deviation of the difference of systolic blood pressure

时间窗: 1 day at visit 2

standard deviation of the difference of systolic blood pressure between new and control device

standard deviation of the difference of diastolic blood pressure

时间窗: 1 day at visit 2

standard deviation of the difference of diastolic blood pressure between new and control device

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Sponsor

研究点 (2)

Loading locations...