Reliability of the Freestyle Libre CGM in the Inpatient Setting During the COVID-19 Surge
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 52
- 试验地点
- 1
- 主要终点
- Comparison of values between two devices
研究概览
简要总结
The FDA has provided an emergency waiver for the use of non-invasive continuous glucose monitors (CGM) in hospitals, so frontline health care providers (in hospitals) can remotely monitor patients with diabetes thus reducing patient's discomfort, limiting exposure to COVID-19 and preserving critical personal protective equipment (PPE). The FreeStyle Libre 14-day system is a continuous glucose monitoring system consisting of a handheld reader (smart phone may be used) and a sensor applied with adhesive to back of the upper arm. In order to evaluate the reliability of the Freestyle Libre CGM for in-patient use, we propose a study which will examine the correlation between the libre CGM data and capillary blood glucose test - the current standard of care taken by the Accuchek Inform II platform.
详细描述
In hospitalized patients diagnosed with COVID-19, diabetes mellitus is associated with a higher mortality. Patients with controlled blood glucose have a lower mortality rate than those who are uncontrolled. However, a rigid blood glucose control - which may lead to hypoglycemia- is associated with higher mortality rates. Thus, adequate glucose monitoring is important for hospitalized diabetic patients.
Previous studies have shown that CGMS, in the outpatient setting, engender better glycemic control thereby reducing risk of hypoglycemia. CGMS offer a replacement for finger stick blood glucose monitoring which may not identify an incidence of hypoglycemia until symptoms drive the patient to test. Furthermore, finger stick blood glucose monitoring is painful and time consuming thus causing many patients to avoid this aspect of diabetes self-care management.
Freestyle Libre CGMs sample interstitial fluid glucose subcutaneously. Measurement frequencies typically range from 1 to 15 minutes and most commonly are every 5 minutes.
Previous studies in hospitalized patients living with diabetes mellitus on basal bolus regimen have shown:
• Use of the CGM increased detection of both hypoglycemic and hyperglycemic events versus those sensed by POCT.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Crossover
- 主要目的
- Other
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 99 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Participants acquired from among all non-critical care in patients who are living with Diabetes Mellitus. Because the patients will have to be instructed on the use of the CGM, this study will be limited to English speaking/reading adults (18 years of age or older), who possess the cognitive and physical ability to participate.
排除标准
- •Minors, pregnant women, and critically ill.
研究组 & 干预措施
Freestyle Libre 14-day CGM
The FreeStyle Libre 14 day system is a continuous glucose monitoring system consisting of a handheld reader and a sensor worn on the back of the upper arm.
干预措施: Freestyle Libre 14 day CGM system (Device)
Freestyle Libre 14-day CGM
The FreeStyle Libre 14 day system is a continuous glucose monitoring system consisting of a handheld reader and a sensor worn on the back of the upper arm.
干预措施: Accuchek Inform II platform (Device)
Accuchek Inform II meter
ACCU-CHEK INFORM II system quantitatively measures glucose in fresh venous, arterial, neonatal heel stick and capillary whole blood from the finger and is used as an aid in monitoring the effectiveness of glucose control
干预措施: Freestyle Libre 14 day CGM system (Device)
Accuchek Inform II meter
ACCU-CHEK INFORM II system quantitatively measures glucose in fresh venous, arterial, neonatal heel stick and capillary whole blood from the finger and is used as an aid in monitoring the effectiveness of glucose control
干预措施: Accuchek Inform II platform (Device)
结局指标
主要结局
Comparison of values between two devices
时间窗: Through end of patient hospitalization, an average of 14 days
Correlation between blood glucose results from the Freestlye Libre 14-day CGM and Accuchek Inform II platform using the Bland-Altman method to determine whether the two glucose methods agree
次要结局