A Clinical Evaluation of the RPS InflammaDry Detector's Sensitivity and Specificity Compared to the Clinical Diagnosis for Confirming Dry Eyes.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 206
- 试验地点
- 8
- 主要终点
- Sensitivity and Specificity of InflammaDry Detector Compared to Clinical Assessment at Confirming a Diagnosis of Dry Eyes.
研究概览
简要总结
The RPS InflammaDry Detector™ is intended to detect elevated MMP-9 in human tears to aid in the diagnosis of patients with signs or symptoms of dry eye disease, in conjunction with other methods of clinical evaluation.
详细描述
The purpose of this clinical trial was to determine the sensitivity and specificity of InflammaDry compared with the clinical assessment of dry eyes.
Patients were screened using clinical history and signs.
Clinical history was performed using the Ocular Surface Disease Index (OSDI) and evaluated for clinical signs:
- positive vital staining of the ocular surface,
- decreased tear breakup time (TBUT),
- reduced corneal sensitivity, and
- decreased functional visual acuity Last, an independent health care professional masked to the clinical evaluation was asked to analyze each InflammaDry test result, independently confirming each result.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- 未提供
排除标准
- 未提供
结局指标
主要结局
Sensitivity and Specificity of InflammaDry Detector Compared to Clinical Assessment at Confirming a Diagnosis of Dry Eyes.
时间窗: 15 minutes
Patients were assessed for signs and Symptoms of Dry Eye using OSDI (Ocular Surface Disease Index), TBUT (Fluorescein Tear Break-up Time), Corneal fluorescein staining and Schirmer Tear Test and compared to the results of the tested device. Sensitivity is the percentage of true positive cases correctly identified by the test, compared to clinical assessment. Specificity is the percentage of true negative cases correctly identified by the test, compared to clinical assessment.