An Evaluation of a New Subcutaneous Glucose Sensor in Pediatric Subjects

Medtronic Diabetes4 个研究点 分布在 1 个国家目标入组 83 人开始时间: 2010年7月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
83
试验地点
4
主要终点
Glucose Sensor Accuracy When Compared to SMBG: Proportion of Glucose Sensor Readings That Met Accuracy Criteria [Time Frame: Days One Through Six of Sensor Use]

研究概览

简要总结

The purpose of this study is to assess performance of a new subcutaneous glucose sensor over a seven day sensor life when used with currently marketed Medtronic Diabetes devices. In addition performance of the new sensor will be calculated for use of the new sensor with proposed new devices using new calibration algorithms.

详细描述

The first generation Medtronic MiniMed Subcutaneous Glucose Sensor (Sof-Sensor) was originally approved by the FDA for commercialization as part of the Continuous Glucose Monitoring System (CGMS) on June 15, 1999 (PMA 980022). The Sensor is composed of a microelectrode with a thin coating of glucose oxidase beneath several layers of biocompatible membrane. This same sensor was used as part of subsequent continuous glucose monitoring (CGM) systems, such as the Guardian REAL-Time and Paradigm REAL-Time sensor augmented insulin pump. The current configuration of sensor has undergone in vitro and in vivo testing. A new second-generation glucose sensor (herein referred to as the Comfort Sensor) has been developed. The first-generation glucose sensor was approved with reported sensor accuracy (MAD) of 18%; it was labeled for maximum use duration of 72 hours, using only the abdomen as an insertion site. The new sensor is shorter and has a smaller diameter, with a smaller gauge introducer needle. The new sensor inserter is designed to be used with the new sensor. The objectives of this study are to 1) Assess performance of the Comfort Sensor when used over a period of seven days with currently available devices, and 2) Assess performance of the Comfort Sensor when used over a period of seven days using new calibration algorithms (post-processed with algorithm for future devices). Accuracy data will be calculated based on comparing calibrated glucose sensor values to glucose meter values during in-home testing. The devices that will be used for gathering sensor data during this study will be: 1) the Comfort Sensor, 2) the MiniLink (Transmitter), and 3) the Guardian REAL-Time Display Device.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
7 Years 至 17 Years(Child)
性别
All
接受健康志愿者
否

入选标准

  • •Males and females aged 7 through 17 inclusive
  • •Diagnosed with Type 1 Diabetes Mellitus
  • •Willingness to perform required study and data collection procedures and to adhere to operating requirements of the Guardian REAL-TIME System
  • •Willingness to perform at least 4 capillary blood glucose tests per day while wearing the Guardian REAL-TIME System
  • •Subject agrees to comply with the study protocol requirements
  • •Informed Consent, Assent, HIPAA Authorization, and California Experimental Subject Bill of Rights (if applicable) signed by the subject and/or parent/guardian
  • •The Subject is willing to wear the Guardian REAL-Time System for 14 days (~340 hours).

排除标准

  • •The Subject has a history of tape allergies that have not been resolved at time of enrollment
  • •The Subject has any skin abnormality (e.g. psoriasis, rash, staphylococcus infection) in the area of sensor placement that have not been resolved at the time of enrollment and would inhibit them from wearing the sensors
  • •Subject has a positive pregnancy test on enrollment
  • •Subject is currently participating in an investigational study (drug or device)

研究组 & 干预措施

Sensor

Experimental

All subjects that wear sensors (all subjects)

干预措施: Sensor wear (Device)

结局指标

主要结局

Glucose Sensor Accuracy When Compared to SMBG: Proportion of Glucose Sensor Readings That Met Accuracy Criteria [Time Frame: Days One Through Six of Sensor Use]

时间窗: Days one through six of sensor use

The primary accuracy parameter (primary effectiveness endpoint) was the comparative readings of paired sensor and SMBG glucose readings, measured on days 1 through 6. Accuracy is defined as within 20% agreement between SMBG and paired sensor (within 20 mg/dL if SMBG ≤ 80 mg/dL). Accuracy ranges from 0 - 100, with higher number suggests better accuracy.

次要结局

  • Device Related Moderate or Device Related Severe Adverse Events(days one through six of sensor wear)