ADRIA - Belos A+ vs DR Clinical Investigation of Arrhythmia Discrimination

Biotronik SE & Co. KG20 个研究点 分布在 2 个国家目标入组 260 人开始时间: 2003年8月1日最近更新:
适应症

试验速览

阶段
4 期
状态
已完成
入组人数
260
试验地点
20
主要终点
Rate of correctly discriminated supraventricular tachyarrhythmia episodes (specificity)

研究概览

简要总结

A unique single-lead dual-chamber implantable cardioverter-defibrillator (ICD) system (produced by Biotronik, Germany) features a conventional electrode in the ventricle (anchored in the ventricular apex) and a floating electrode (ring on the lead body) in the atrium, capable of sensing atrial electrical signals. The purpose of this study is to determine whether this system detects a supraventricular tachyarrhythmia (e.g. atrial fibrillation, atrial tachycardia) in the equivalent manner as conventional dual-lead dual-chamber ICDs using two separate electrodes anchored in the ventricle and in the atrium, respectively.

详细描述

Supraventricular tachyarrhythmia (SVT) is the main cause of inappropriate therapy in patients with single-chamber implantable cardioverter-defibrillators (ICDs). To minimize inappropriate shock delivery, ICDs should sense both atrial and ventricular intracardiac signals. Dual-chamber ICDs are used for this purpose, but are associated with increased postoperative complications due to the implantation of a separate atrial lead. It has to be shown that a novel single-lead dual-chamber ICD-system with enhanced SVT discrimination can achieve the same specificity in discriminating SVT episodes as conventional dual-chamber ICD, without the disadvantage of the implantation of several leads. In this study patients eligible for dual-chamber ICD therapy who do not need atrial pacing will receive either a single-lead dual-chamber ICD (Belos A+ and Kainox A+ electrode) or a dual-lead dual-chamber ICD (Belos DR). SMART detection algorithm will be used in both study groups for discrimination between atrial and ventricular tachyarrhythmias. Atrial tachyarrhythmia episodes are facultatively induced in both groups at implantation or predischarge via a bipolar stimulation catheter. Induced episodes and the corresponding ICD intervention (detection and therapy or inhibition of therapy) are documented. Follow-ups are scheduled for 1, 3, 6 and 12 months after implantation. ICD intervention, particularly related to spontaneous SVT episodes, will be evaluated based on ICD diagnostic memory data interrogated at follow-up controls.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • ICD indication according to American College of Cardiology/American Heart Association (ACC/AHA) guidelines.
  • Informed consent

排除标准

  • Permanent atrial fibrillation
  • Requirement for atrial pacing
  • Patient is underage
  • No signed patient agreement
  • Patient life expectancy under 6 months

结局指标

主要结局

Rate of correctly discriminated supraventricular tachyarrhythmia episodes (specificity)

次要结局

  • Complication rate (e.g. lead dislocation, lead fracture)
  • Duration of implantation
  • Electrode performance (P-/R-wave amplitudes, ventricular pacing threshold and impedance)
  • Sensitivity (rate of correctly discriminated episodes of ventricular fibrillation or ventricular tachycardia)