Sensitivity and Specificity of the New Mindray 3-Directional Neuromuscular Transmission Module Versus the Relaxometer Mechanomyograph and Versus TOF-Watch Acceleromyograph for Neuromuscular Block Monitoring
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 48
- 试验地点
- 2
- 主要终点
- Pharmacodynamics parameters (onset time 95% maximum depression, clinical duration T1 25%, and full recovery 0.8-0.9 TOF ratio neuromuscular block
研究概览
简要总结
A new NMT module from Mindray (Mindray Co. Shenzhen, People's Republic of China.) claims to measure 3 directional accelerography. The aim of the study is to compare the neuromuscular block of rocuronium 0.6 mg/kg (twice the 95% effective dose, ED95) monitored by the NMT versus that monitored by the Relaxometer Mechanomyograph on the other hand in Group 1, and versus the TOF-Watch on the other hand in Group 2 to clinically evaluate the new system for its diagnostic accuracy.
详细描述
Mechanomyography (MMG) is regarded as the standard method for precise quantification of neuromuscular block. The conventional MMG measures the exact force of muscle contraction in response to electric stimulation of the ulnar nerve. MMG quantifies the neuromuscular function by measurement of the force displacement. However the equipment is rather bulky, takes time to set up, requires a rigid support of the arm in an often-crowded operating room. This limits its clinical use in the daily anesthesia practice. Acceleromyography is based on the acceleration of a piezo-electrode. A new NMT module from (Shenzhen, People's Republic of China.) claims to measure 3 directional accelerography. The aim of the study is to compare the neuromuscular block of rocuronium 0.6 mg/kg (twice the 95% effective dose, ED95) monitored by the NMT to that monitored by the Relaxometer Mechanomyograph in Group 1, and versus TOF-Watch Acceleromyograph in Group 2 to clinically evaluate the new system for its diagnostic accuracy.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •24 patients with Mindray NMT monitoring on one hand versus MMG Mechanomyograph on the other hand in Group 1 with both hands accessible in supine position alternately right and left hands monitoring neuromuscular block of rocuronium neuromuscular blocking agent 0.6 mg/kg
- •24 patients with Mindray NMT monitoring one hand versus TOF-Watch Acceleromyograph on the other hand in Group 2 with both hands accessible in supine position alternately right and left hands monitoring neuromuscular block of rocuronium neuromuscular blocking agent 0.6 mg/kg.
排除标准
- •Both hand inaccessible in supine position, patients receiving drugs interfering with rocuronium neuromuscular block, or patients with small joint arthritis that could affect neuromuscular monitoring.
结局指标
主要结局
Pharmacodynamics parameters (onset time 95% maximum depression, clinical duration T1 25%, and full recovery 0.8-0.9 TOF ratio neuromuscular block
时间窗: one year
this will determine the Sensitivity and specificity versus MMG mechanomyography
次要结局
- Bland and Altman bias and limits of agreement(one year)
研究者
Ashraf Dahaba, MD
Principal Investigator
Suez Canal University