The Accuracy of the Dormotech VLAB During Stable Hypoxia Plateaus

Dormotech Medical1 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2024年5月15日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
12
试验地点
1
主要终点
Measurement of arterial HbO2 saturations from 100% down to 70%

研究概览

简要总结

The purpose of the study is to determine the accuracy of the Dormotech VLAB (SpO2) Sensor in accordance with the FDA Guidance on Pulse Oximeters - Premarket Notification Submissions [510(k)s], in range of arterial HbO2 saturations from 100% down to 70%.

This was a prospective study with healthy volunteers.

The primary objective of the study was to determine the accuracy of the Dormotech VLAB sensor from saturations from 100 down to 70%. As the primary endpoint, the accuracy of the device during non-motion tests was determined by comparing the pulse oximeter reading during brief, steady-state hypoxia plateaus with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a multi-wavelength hemoximeter).

详细描述

The purpose of the study is to determine the accuracy of the Dormotech VLAB (SpO2) Sensor in accordance with the FDA Guidance on Pulse Oximeters - Premarket Notification Submissions [510(k)s], in the range of arterial HbO2 saturations from 100% down to 70%.

The subjects represent a diverse group in terms of age, height, weight, ethnicity, and skin types, which is beneficial for generalizing findings across different demographics. Subjects span all six types, indicating a wide range of skin types from very fair (Type I) to deeply pigmented dark (Type VI).

This was a prospective study with healthy volunteers.

The primary objective of the study was to determine the accuracy of the Dormotech VLAB sensor from saturations from 100 down to 70%. As the primary endpoint, the accuracy of the device during non-motion tests was determined by comparing the pulse oximeter reading during brief, steady-state hypoxia plateaus with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a multi-wavelength hemoximeter).

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
18 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Subject is male or Subject is male or female, aged 18 and <
  • Subject is in good health with no evidence of any medical problems.
  • Subject is fluent in both written and spoken English.
  • Subject has provided informed consent and is willing to comply with the study procedure

排除标准

  • Subject is obese (BMI >35).
  • Subject has known history of heart disease, lung disease, kidney or liver disease.
  • Diagnosis of asthma, sleep apnea, or use of CPAP.
  • Subject has diabetes.
  • Subject has a clotting disorder.
  • Subject uses any anticoagulative medication (blood thinner).
  • Subject has a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
  • Subject has any other serious systemic illness.
  • Subject is a current smoker.
  • Subject with any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators would interfere with the sensors working correctly.
  • Subject has a history of fainting or vasovagal response.
  • Subject has a history of sensitivity to local anesthesia
  • Subject has a diagnosis of Raynaud's disease
  • Subject has claustrophobia
  • Subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
  • Subject is pregnant, lactating or trying to get pregnant.
  • Subject is unable or unwilling to provide informed consent or is unable or unwilling to comply with study procedures.
  • Subject has any other condition, which in the opinion of the investigators would make them unsuitable for the study.

结局指标

主要结局

Measurement of arterial HbO2 saturations from 100% down to 70%

时间窗: day 1

At the start of each testing cycle, one blood sample was taken while breathing room air. Subjects then breathed through a partial rebreathing circuit with a nose clip, a mixture of nitrogen, air and carbon dioxide to reach six different stable levels (plateaus) of oxyhemoglobin saturation between 70-100%.