Validity (and Reliability) of Two Forms of an Accelerometer-Based Intravaginal Device for Detecting Pelvic Floor Motion
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 30
- 试验地点
- 2
- 主要终点
- Validate leva measurements with transperineal ultrasound (TPUS)
研究概览
简要总结
The leva and PFDx devices accurately reflect pelvic floor motion consistent with TPUS evaluation. During a voluntary PFM contraction, the change in angle from a position of rest to maximal effort that is detected by each device is positively correlated with TPUS measurements and Modified Oxford scores.
详细描述
Thirty women 18 years or older will be recruited from OB/GYN, FPMRS or other clinics. Following informed consent and successful screening, subjects will undergo testing as outlined below. Five subjects (all will be invited until five subjects have agreed) will return to clinic in 1-2 weeks to undergo repeat testing using all modalities.
After obtaining informed consent, subject information will be collected, including demographic data, BMI, and medical history (including ob/gyn history). Subjects will complete urinalysis for pregnancy. Subjects will also complete the Pelvic Floor Distress Inventory (PFDI-20) questionnaire. All subjects will undergo a complete pelvic examination including pelvic organ prolapse quantification (POP-Q), a digital assessment of PFM strength (Modified Oxford Scale) and receive a TPUS. Subjects will also complete a series of PFM contractions using the leva and PFDx devices to measure angle changes, and undergo vaginal assessment with a perineomter.
Pelvic floor muscle assessment Prior to recording device measurements, a correct pelvic floor muscle contraction will be verified via manual pelvic floor assessment. Subjects will be instructed to perform a "lift-and-squeeze" maneuver of the pelvic floor, and correct PFM contraction will be verified by the primary examiner (LHQ). PFM strength will then be assessed via digital palpation by the examiner inserting the first and/or second fingers approximately 4 cm into the vagina and palpating the right and left PFM during contraction. Strength will be graded according to the 6-point Modified Oxford Scale (0=no contraction; 1='flicker'; 2=weak; 3=moderate; 4=good; 5=strong). The mean score of the right and left sides of the vagina will be calculated and used for correlation analysis.
Perineometry PFM strength will be measured using a Peritron™ perineometer, which is a 28mm diameter compressible probe connected to a handheld microprocessor. It provides readings of pelvic floor contraction in cm of H20 with precision up to one decimal point, and it has been demonstrated to show reproducible and reliable measurements. The resting vaginal pressure and two peak pressures will be measured. For this study, PFM strength is defined as the augmented vaginal pressure, calculated by measuring the average of 2 peak pressures minus the resting vaginal pressure.
TPUS measurements TPUS imaging will be completed at the time of the study visit using the BK Medical Ultrafocus (Peabody, MA) and 8802 transperineal transducer. The probe will be covered with ultrasound gel and a probe cover. All ultrasound studies will be performed in the office setting with the patient in the dorsal lithotomy position, with hips flexed and abducted. No preparation is required, and no rectal or vaginal contrast will be used. Patients will be instructed to arrive at the office with a partially full bladder and to avoid excessive pressure on surrounding structures that might distort the anatomy. In the sagittal view, dynamic ultrasound parameters will be measured relative to a horizontal reference line at the level of the infero-posterior margin of the pubic symphysis. Dynamic images will be obtained during PFM contractions and at rest. At least 3 cycles of PFM contraction and relaxation will be recorded for each individual.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Health Services Research
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 100 Years(Adult, Older Adult)
- 性别
- Female
- 接受健康志愿者
- 是
入选标准
- •18 years of age or older
- •Presence of a vagina
- •Ability to read and write English
- •Reports they are able to perform a kegel exercise
排除标准
- •Inability to perform a kegel exercise
- •Pelvic organ prolapse greater than ICS POP-Q Stage II
- •Known, untreated pelvic, vaginal or urinary tract infection
- •Currently pregnant
研究组 & 干预措施
Device usage
All participants will complete a survey and undergo an evaluation to test the movement of their pelvic floor with the PFDx device and leva device
干预措施: PFDx device and leva device (Device)
结局指标
主要结局
Validate leva measurements with transperineal ultrasound (TPUS)
时间窗: Through final study visit (about 3 months)
During a voluntary pelvic floor muscle contraction, the change in angle from a position of rest to maximal effort that is detected by the leva device will be compared to the change in distance from the pubic bone to the levator plate measured by 2D TPUS during the same task.
次要结局
- Determine the test-retest reliability of perineometry angle measurements over 2 time points(Through final study visit (about 3 months))
- Determine the test-retest reliability of PFDx angle measurements over 2 time points(Through final study visit (about 3 months))
- Compare leva angle change measurements with Modified Oxford, Brinks(Through final study visit (about 3 months))
- Compare PFDx angle change measurements with Modified Oxford, Brinks(Through final study visit (about 3 months))
- Compare PFDx to a perineometer(Through final study visit (about 3 months))
- Compare leva to a perineometer(Through final study visit (about 3 months))
- Determine the test-retest reliability of Manual muscle testing angle measurements over 2 time points(Through final study visit (about 3 months))
- Determine the test-retest reliability of ultrasound angle measurements over 2 time points(Through final study visit (about 3 months))
- Determine the test-retest reliability of leva angle measurements over 2 time points(Through final study visit (about 3 months))