Validation Study of WITHINGS BPM Core for the Detection of Atrial Fibrillation

Withings2 个研究点 分布在 1 个国家目标入组 160 人开始时间: 2020年7月6日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
进行中(未招募)
发起方
Withings
入组人数
160
试验地点
2
主要终点
Sensitivity of the automatic classification in AF by BPM Core single-lead ECG against reference 12-lead ECG

研究概览

简要总结

The proposed clinical study aims to validate the diagnostic performance, compared to a reference ECG, of the electrocardiographic function of the BPM Core developed by Withings for the automatic identification of atrial fibrillation (AF).

详细描述

BPM Core is a blood-pressure monitor with an build-in single-channel electrocardiogram similar to a lead I.

Primary objective:

The primary objective is to demonstrate that Withings BPM Core has acceptable sensitivity and specificity in the automatic identification of AF and SR in comparison with a gold-standard diagnosis from a 12-lead ECG by board certified cardiologists.

The algorithm classifies the device signals into four categories: normal sinus rhythm (RSN), AF, arrhythmia other than AF, or non-interpretable. The latter category includes signals of insufficient quality to be interpreted with confidence.

This classification will be compared to the diagnosis made by three independent, blinded cardiologists based on a 12-lead ECG. The final diagnosis retained will be decided by majority with adjudication by a consensus of 3 to 6 cardiologists. The sensitivity and specificity of the device will then be estimated.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 100 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • male or female who are 18 years or older
  • Subject who signed the written informed consent form,
  • Able to communicate effectively with and follow instructions from the study staff
  • Person with an arm circumference between 22 and 42 cm

排除标准

  • Vulnerable subject with regard to regulations
  • Pregnant, parturient or breastfeeding woman,
  • Subject who is deprived of liberty by judicial, medical or administrative decision,
  • Underage subject,
  • Legally protected subject, or subject who is unable to sign the written informed consent form,
  • Subject who is not beneficiary or not affiliated to a social security scheme,
  • Subject within several of the above categories,
  • Subject who refused to participate in the study,
  • Subject in linguistic or psychic incapacity to sign a written informed consent form,
  • Subject in physical incapacity to wear a blood pressure monitor
  • Subject with electrical stimulation by pacemaker
  • Significant tremor that prevents subject from being able to hold still
  • Known sensitivity to medical adhesives, isopropyl alcohol, or ECG electrodes

研究组 & 干预措施

atrial fibrillation (AF)

Patients diagnosed with AF during reference ECG

干预措施: ECG measurement (Diagnostic Test)

sinus rhythm (SR)

Patients diagnosed with SR during reference ECG

干预措施: ECG measurement (Diagnostic Test)

结局指标

主要结局

Sensitivity of the automatic classification in AF by BPM Core single-lead ECG against reference 12-lead ECG

时间窗: 1 year

sensitivity from the 2x2 confusion matrix AF vs 'non-AF'

次要结局

  • Sensitivity of the cardiologists' classification in SR from the BPM Core ECG(1 year)
  • Accuracy of the positivity of P-waves of single-lead ECG from BPM Core(1 year)
  • Sensitivity of the cardiologists' classification in AF from the BPM Core ECG(1 year)
  • Accuracy of the visibility of QRS-complexes of single-lead ECG from BPM Core(1 year)
  • Accuracy of the positivity of QRS-complexes of single-lead ECG from BPM Core(1 year)
  • Accuracy of the visibility of T-waves of single-lead ECG from BPM Core(1 year)
  • Accuracy of the visibility of P-waves of single-lead ECG from BPM Core(1 year)