Evaluating Active Esophageal Cooling During Cardiac Ablation Procedures

Advanced Cooling Therapy, Inc., d/b/a Attune Medical2 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2019年6月20日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
20
试验地点
2
主要终点
Total Time of Active Ablation Procedure

研究概览

简要总结

Left atrial catheter ablation including pulmonary vein isolation is a standard therapy in the management of symptomatic atrial fibrillation; however thermal esophageal injury is a known potential consequence of this procedure. Delivery of radiofrequency (RF) energy necessary to perform left atrial ablation has the potential to cause injury to the nearby esophagus including ulceration, hematoma, spasm, esophageal motility disorders, and, in the most extreme case, atrial-esophageal fistula (AEF). Esophageal mucosal lesions are the likely precursor to AEF, and esophageal mucosal lesions have been detected on post-ablation endoscopy after pulmonary vein isolation with an incidence ranging from 3% to 60%.

Active esophageal cooling during RF ablation as a means of esophageal injury prevention has been investigated through mathematical models, pre-clinical studies, and in clinical trials. Existing data support the efficacy of this approach, but the practice has not been widely adopted due to lack of a commercially available device.

The aim or purpose of this study is to evaluate the impact on procedural efficiency of ablation procedures performed using esophageal heat transfer to cool the esophagus during left atrial RF ablation.

详细描述

This study is a prospective, pilot study using the Attune Medical esophageal heat transfer device to actively cool the esophagus during RF ablation procedures. This design is appropriate to gather the data needed regarding overall procedural time, compare this to historical controls, and estimate a sample size for a larger study powered for statistical significance.

The subjects will undergo preparation and anesthesia procedures following standard practice. Once the patient is intubated, the esophageal heat transfer device will be placed into the esophagus according to the Instructions for Use (IFU). The device will remain in place until the ablation procedure is completed and will be removed before extubation. Posterior left atrial wall ablation using standard parameters will only be performed when the esophageal heat transfer device has reached a temperature of 4-6 degrees C for at least 2 minutes. The device will be set to neutral or warming temperature (37-42 degrees C) during other aspects of the procedure (such as mapping and anterior wall ablations).

All patients will be followed up in total for 6 weeks (Long Term Follow-up visit) after the procedure to document any clinical complication related to thermal esophageal injuries if applicable.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Prevention
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adult patients (age over 18 years)
  • Undergoing first left atrial ablation for the treatment of atrial fibrillation (AF) including pulmonary vein isolation
  • Undergoing catheter-based ablation procedure using radiofrequency energy
  • Patients must be able to understand and critically review the informed consent form.
  • Subjects must understand and agree to study requirements and sign a written informed consent.

排除标准

  • Patients who are unable to provide informed consent.
  • History of prior atrial fibrillation (AF) ablation procedures.
  • Significant co-morbidities that preclude standard ablation procedure.
  • Patients with <40 kg of body mass
  • Patients with relevant esophageal pathology (e.g. esophageal cancer)

结局指标

主要结局

Total Time of Active Ablation Procedure

时间窗: Study Day 1 for all patients enrolled, during left atrial ablation procedures using radiofrequency energy, up to 5 hours

Study Day 1 for all patients enrolled, during left atrial ablation procedures using radiofrequency energy, up to 5 hours. Measured as total procedure time as documented by clinician recorder and research coordinator.

次要结局

  • Total Procedure Time(Study Day 1 for all patients enrolled from patient entry to Electrophysiology (EP) lab until post procedure discharge to Post-Anesthesia Care Unit (PACU))
  • Number of Procedural Pauses During Left Atrial Instrumentation(Study Day 1 for all patients enrolled, during left atrial ablation procedures using radiofrequency energy, up to 5 hours)