System accuracy evaluation of 7 blood glucose monitoring systems based on ISO 15197:2013 / EN ISO 15197:2015 - IfDT-1948-I

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm (IfDT)0 个研究点目标入组 124 人开始时间: 2020年11月9日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
发起方
入组人数
124

研究概览

简要总结

According to ISO 15197:2013 and EN ISO 15197:2015, at least three reagent system lots of a blood glucose monitoring system have to be tested. For each of these three lots, at least 95% of measurement results have to lie within ±15 mg/dl or ±15% of the results obtained with manufacturer’s reference method for reference results <100 mg/dl or =100 mg/dl, respectively. In addition, at least 99% of results across all three lots have to fall within CEG regions A and B (CEG: Consensus Error Grid). In the primary analysis versus the manufacturer’s reference method (for all systems YSI 2300 STAT Plus, glucose oxidase-based), the following results were found (n = 200 measurement results each): - System A: 100.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System B: 95.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System C: 96.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System D: 87.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System E: 95.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System F: 66.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System G: 89.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B In the secondary analysis versus the alternative comparator (for all systems Cobas Integra 400 plus, hexokinase-based), the following results were found (n = 200 measurement results each): - System A: 98.5% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System B: 91.5% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System C: 88.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System D: 93.5% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System E: 89.5% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System F: 42.0% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B - System G: 77.5% within ±15 mg/dl or ±15%, 100.0% within CEG regions A and B

研究设计

研究类型
Interventional
分配方式
N/a: Single Arm Study
盲法
Open (masking not used)

入排标准

年龄范围
18 Years 至 one(—)
性别
All

入选标准

  • Signed informed consent form
  • - Subjects are legally competent and capable to understand character, meaning and consequences of the study.
  • - If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
  • - Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
  • - Signature of subjects to document consent with these procedures on informed consent form.

排除标准

  • - Pregnancy or lactation period if venous samples are needed
  • - Severe acute disease that compromises the subject’s capability to participate in the study (at the study physician’s discretion)
  • - Severe chronic disease with potential risk during the test procedures (at the study physician’s discretion)
  • - Medical relevant anemia
  • - Suspected lack of compliance
  • - Dependence on investigator or sponsor
  • - If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy:
  • - Pregnancy or lactation period
  • - Subjects with type 1 diabetes, suffering from
  • - Coronary heart disease
  • - Condition after myocardial infarction
  • - Condition after cerebral events
  • - Peripheral arterial occlusive disease
  • - Hypoglycemia unawareness