临床试验/NCT05895682
NCT05895682已完成不适用

A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers

Seers Technology Co., Ltd.2 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2023年4月29日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
12
试验地点
2
主要终点
Accuracy of SpO2

研究概览

简要总结

It is important to evaluate the accuracy of the pulse oximeter, which is commonly used in clinical practice. The US FDA only allows the use of devices that meet the ISO 80601-2-61:2017 standard in the United States. According to the ISO 80601-2-61:2017 standard, the accuracy of the oxygen saturation (SpO2) of the pulse oximeter should have an error range of less than 4.0% compared to the actual arterial blood oxygen saturation (SaO2) in the range of 70-100%. To prove compliance with this requirement, it is necessary to conduct a clinical trial that induces hypoxemia in healthy adults and compares and evaluates the arterial blood oxygen saturation values of the pulse oximeter's SpO2 and the arterial blood oxygen saturation values of the carbon monoxide-oxygen meter (CO-oximeter).

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
20 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • healthy adults between 20 and 50 years
  • volunteers who have plasma proportion of carboxyhemoglobin 3%
  • volunteers who have plasma proportion of methemoglobin 2%
  • volunteers who have total plasma hemoglobin concentration over 10 g/dL

排除标准

  • History of respiratory diseases
  • History of cardiovascular disease
  • Smokers (including ex-smokers)
  • History of fainting
  • History of diabetes
  • Obesity (BMI> 30 kg/m2)
  • Allen's test results, those who do not recover the color of their fingers and palms within 10 seconds
  • History of allergic reaction to lidocaine, a local anesthetic

结局指标

主要结局

Accuracy of SpO2

时间窗: 2 day

The root-mean-square of the difference between the observed pulse oximetry value (SpO2) and the carbon monoxide-oxygenometer arterial blood reference value (SRi) within the range of 70-100% of arterial oxygen saturation

次要结局

未报告次要终点