JPRN-UMIN000037436已完成未知
Validation of Ankle-Barchial Index Measurement Using a Multi-Cuff Oscillometric Device Versus the Standard Doppler Method. - Validation of Ankle-Barchial Index Measurement Using a Multi-Cuff Oscillometric Device
ara Medical University0 个研究点目标入组 120 人开始时间: 2019年8月1日最近更新:
适应症
试验速览
- 阶段
- 未知
- 状态
- 已完成
- 发起方
- 入组人数
- 120
研究概览
简要总结
In total, 118 patients were evaluated. The prevalence of Doppler ABI 0.90 was 45.8%. Taking the Dop-ABI as the reference, the sensitivity, specificity, accuracy, positive and negative predictive values of oscillometric ABI during the first measurement by the first observer were 89.1%, 94.4%, 94.1%, 91.8%, and 92.4%, respectively.
研究设计
- 研究类型
- Observational
入排标准
- 年龄范围
- ot applicable 至 ot applicable(—)
- 性别
- All
入选标准
- 未提供
排除标准
- •Cardiac arrhythmia: Atrial fibrillation, atrial flutter, frequent supra and ventricular ectopic beats Patients under dialysis
研究者
相似试验
Unknown
不适用
Comparison of 3 Ultrasonic Techniques for Measuring the Systolic Pressure IndexAtherosclerosis ObliteransAtherosclerosis Right LegAtherosclerosis Left LegPeripheral Arterial DiseaseNCT04440072University Hospital, Caen150
已完成
不适用
Brachial Artery Pulsatility Index in Supraclavicular BlockBrachial Plexus AnesthesiaDiagnostic AccuracyNCT06475638Cairo University89
已完成
不适用
Validation of WatchBP Office Ankle-brachial Index (ABI) FunctionHypertensionNCT01002989University of Athens98
Unknown
不适用
The Prevalence of a Low Ankle-Brachial Index in Acute Cerebral Ischemia.Ischemic StrokeNCT03948399Jan Biziel University Hospital No 2 in Bydgoszcz200
已完成
不适用
Comparison of the Ankle-Brachial Index Measurement Using a Specific Oscillometric Device vs. the Doppler MethodPeripheral Arterial Disease, Ankle Brachial Index, DopplerNCT03511612University Hospital, Limoges51