临床试验/NCT04548986
NCT04548986已完成不适用

Single-center Prospective Study to Validate the Performance of the Aktiia Optical Blood Pressure Monitoring (OBPM) Device at the Wrist Against Ambulatory Blood Pressure Monitoring (ABPM)

Cyril Pellaton2 个研究点 分布在 1 个国家目标入组 63 人开始时间: 2020年8月6日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
63
试验地点
2
主要终点
Mean error for blood pressure

研究概览

简要总结

This study aims to assess the performances of optical blood pressure monitoring device, Aktiia.product in the context of cardiac rehabilitation program.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Screening
盲法
None

入排标准

年龄范围
21 Years 至 65 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adult subjects (aged between 21 and 65)
  • Subjects fluent in written and spoken French
  • Subjects enrolled in a 12-weeks cardiac rehabilitation program
  • Subjects agreeing to participate
  • Subjects that have signed the informed consent form

排除标准

  • Subjects with tachycardia (heart rate at rest > 120bpm)
  • Subjects with atrial fibrillation
  • Subjects with severe heart failure (LVEF<35%)
  • Subjects with severe renal dysfunctions (eGFR < 30mL/min/1.73 m2)
  • Subjects with pheochromocytoma
  • Subjects with Raynaud's disease
  • Subjects with trembling and shivering
  • Subjects with interarm systolic difference > 15 mmHg
  • Subjects with interarm diastolic difference > 10 mmHg
  • Subjects with arm paralysis
  • Women in known pregnancy
  • Subjects with an arteriovenous fistula
  • Subjects with arm amputations
  • Subjects with the upper arm circumference > 64 cm
  • Subjects with the wrist circumference > 22 cm
  • Subjects with the exfoliative skin diseases
  • Subjects with lymphoedema

结局指标

主要结局

Mean error for blood pressure

时间窗: up to 24 hours

The mean error between the Reference and Aktiia.product for SYS and DIA in mmHg

Mean absolute error for blood pressure

时间窗: up to 24 hours

The mean absolute error between the Reference (standard ambulatory blood pressure monitor) and Aktiia.product for systolic (SYS) and diastolic (DIA) blood pressure in mmHg

Standard deviation of the error for blood pressure

时间窗: up to 24 hours

The standard deviation of the error between Reference and Aktiia.product for SYS and DIA in mmHg

Mean absolute error for heart rate

时间窗: up to 24 hours

The mean absolute error between the Reference (standard ambulatory blood pressure monitor) and Aktiia.product for heart rate in bpm

Mean error for heart rate

时间窗: up to 24 hours

The mean error between the Reference and Aktiia.product for heart rate in bpm

Standard deviation of the error for heart rate

时间窗: up to 24 hours

The standard deviation of the error between Reference and Aktiia.product for heart rate in bpm

次要结局

  • Assessment of systolic and diastolic blood pressure control during a 12 weeks cardiac rehabilitation program(12 weeks)
  • Evaluation of the impact of ABPM induced stress on BP readings using inferential statistics.(1 week)
  • Evaluation of patient's satisfaction for the two devices through a survey with multiple questions(up to 24 hours)
  • Assessment of the calibration stability of Aktiia OBPM algorithms(12 weeks)

研究者

申办方类型
Other
责任方
Sponsor Investigator
主要研究者

Cyril Pellaton

Cyril Pellaton

Hopital Neuchatelois

研究点 (2)

Loading locations...

相似试验