Non-Contact Air Tonometry (Topcon CT-1®) Versus Goldmann Applanation Tonometry: a Device Comparison Study

University of Zurich1 个研究点 分布在 1 个国家目标入组 58 人开始时间: 2017年4月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
58
试验地点
1
主要终点
Comparison of GAT with air tonometry Topcon CT1

研究概览

简要总结

In this prospective device comparison study with 58 healthy eyes, the Topcon CT-1® showed statistically significant elevated intraocular pressure measurements compared to the gold standard Goldmann application tonometry.

详细描述

Introduction

Intraocular pressure (IOP) measurement is crucial in glaucoma diagnosis and treatment. We compared a non-contact air tonometer (Topcon CT-1®) with the gold standard Goldmann Applikation tonometer (GAT).

Methods:

In this prospective, mono centric device comparison study, only healthy eyes were enrolled. IOP was measured three times with the Topcon CT-1® followed by three times GAT. Mean IOP was calculated and compared with a paired t-test, a linear regression and a Bland-Altman analysis.

研究设计

研究类型
Observational
观察模型
Case Control
时间视角
Prospective

入排标准

年龄范围
18 Years 至 99 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Healthy volunteers who agreed to be included in the study and were able to comply with the measurement procedures
  • Exclusion criteria were cor-neal abnormalities (such as oedema, scars or dystrophies, astigmatism,hyperopia or myopia > 3 dioptres), a history of ocular surgery or recent(<47 weeks) ocular disorder.

排除标准

  • 未提供

结局指标

主要结局

Comparison of GAT with air tonometry Topcon CT1

时间窗: 7 months

Difference in mmHg between these two devices

次要结局

未报告次要终点