Non-Contact Air Tonometry (Topcon CT-1®) Versus Goldmann Applanation Tonometry: a Device Comparison Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 58
- 试验地点
- 1
- 主要终点
- Comparison of GAT with air tonometry Topcon CT1
研究概览
简要总结
In this prospective device comparison study with 58 healthy eyes, the Topcon CT-1® showed statistically significant elevated intraocular pressure measurements compared to the gold standard Goldmann application tonometry.
详细描述
Introduction
Intraocular pressure (IOP) measurement is crucial in glaucoma diagnosis and treatment. We compared a non-contact air tonometer (Topcon CT-1®) with the gold standard Goldmann Applikation tonometer (GAT).
Methods:
In this prospective, mono centric device comparison study, only healthy eyes were enrolled. IOP was measured three times with the Topcon CT-1® followed by three times GAT. Mean IOP was calculated and compared with a paired t-test, a linear regression and a Bland-Altman analysis.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Control
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 99 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Healthy volunteers who agreed to be included in the study and were able to comply with the measurement procedures
- •Exclusion criteria were cor-neal abnormalities (such as oedema, scars or dystrophies, astigmatism,hyperopia or myopia > 3 dioptres), a history of ocular surgery or recent(<47 weeks) ocular disorder.
排除标准
- 未提供
结局指标
主要结局
Comparison of GAT with air tonometry Topcon CT1
时间窗: 7 months
Difference in mmHg between these two devices
次要结局
未报告次要终点