跳至主要内容
临床试验/NCT05946538
NCT05946538尚未招募不适用

Clinical Evaluation of SARS-COV-2, Influenza and RSV 8-Well MT-PCR Panel for In Vitro Diagnostics

AusDiagnostics Pty Ltd.0 个研究点目标入组 1,000 人开始时间: 2023年7月最近更新:
适应症

试验速览

阶段
不适用
状态
尚未招募
入组人数
1,000
主要终点
Number of participants with positive samples as evaluated by the AusDiagnostics SARS-COV-2, Influenza and RSV 8-Well MT-PCR Panel and the BioFire Respiratory Panel 2.1.

研究概览

简要总结

This study is a prospective accuracy observational study to evaluate the performance of the Highplex System SARS-Cov-2, Influenza and RSV 8-well in vitro diagnostic medical device panel and generate performance data. The precision of results will be evaluated in concordance with the comparator predicate device panel BioFire Respiratory Panel 2.1 (RP2.1).

详细描述

This study is a prospective accuracy observational study to evaluate the performance of the Highplex System SARS-Cov-2, Influenza and RSV 8-well in vitro diagnostic medical device panel and generate performance data. The precision of results will be evaluated in concordance with the comparator predicate device panel BioFire Respiratory Panel 2.1 (RP2.1). This study will utilise prospective sample respiratory swabs from individuals with paired test data, presenting with influenza-like symptoms (e.g., cough, nasal congestion, rhinorrhoea, sore throat, fever, headache, myalgia), as well as asymptomatic individuals. To meet the target number of 400 positive samples (50 samples per target; 8 targets in total), a required sampling size of 1000 participants will need to be screened to account for unviable samples and screen failures. Prospective respiratory samples will include nasopharyngeal swabs as described in Intended Use of the product.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

性别
All
接受健康志愿者

入选标准

  • The sample is derived from an individual who has the ability to provide consent or from a legally acceptable representative, parent(s), or legal guardian and have consented for their sample and demographic information to be taken
  • The sample is a nasopharyngeal swab
  • The sample is collected in universal transport media (UTM) or viral transport media (VTM), using FDA approved swabs
  • The sample is collected from an individual who has a presumptive diagnosis or is presenting with influenza-like symptoms (e.g., cough, nasal congestion, rhinorrhoea, sore throat, fever, headache, myalgia) or is suspected to be asymptomatic.

排除标准

  • Sample does not meet acceptable specimen criteria in that is not a nasopharyngeal swab
  • The sample is collected using swabs with wood shafts or other materials known to inhibit growth of influenza viruses

结局指标

主要结局

Number of participants with positive samples as evaluated by the AusDiagnostics SARS-COV-2, Influenza and RSV 8-Well MT-PCR Panel and the BioFire Respiratory Panel 2.1.

时间窗: 6 months

The study will be considered to have reached its primary endpoint when performance data from 50 positive samples for each panel target has been collected and evaluated for precision and concordance with the comparator predicate device panel BioFire Respiratory Panel 2.1 (RP2.1).

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor

相似试验