Protocol for the Evaluation of Dario™ Blood Glucose Monitor (BGMS) Performance at Elevated Altitude
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 52
- 试验地点
- 2
- 主要终点
- Dario Blood Glucose Monitoring System Accuracy at High Altitude (ISO 15197:2003)
研究概览
简要总结
The purpose of this study is to assess the performance of the Dario Blood Glucose Monitor (BGMS), at elevated altitudes, by obtaining whole blood fingerstick samples from individuals and analyzing the blood on the Dario Blood Glucose Monitor (BGMS) and comparing the readings to readings obtained from the YSI 2300 STAT Plus glucose analyzer.
详细描述
The study will be performed with 50 fresh capillary blood samples, collected by fingerstick using Dario single use lancet. Additional fingerstick with dedicated disposable lancets will be used to collect sufficient capillary fresh blood to be analyzed by the YSI 2300 STAT Plus analyzer. Additional fingersticks may be necessary if volume of blood from first fingerstick is insufficient.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Single Group
- 主要目的
- Other
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Age of subject is 18 years or older
- •Able to speak and read English
排除标准
- •Has any condition that the principal investigator believes may interfere in the subject's participation in the trial.
结局指标
主要结局
Dario Blood Glucose Monitoring System Accuracy at High Altitude (ISO 15197:2003)
时间窗: 6 seconds
Accuracy of Dario Blood Glucose Monitoring System when compared to Yellow Springs Instrument 2300 reference method at 10,152 feet above sea level. Acceptance criteria set per (International Organization for Standardization) ISO 15197:2003. Greater than or equal to 95% of the Dario results shall fall within plus or minus 15 mg/dL of the reference method value (for glucose concentrations \<75 mg/dL). For glucose concentrations greater than or equal to 75 mg/dL, the Dario results shall fall within plus or minus 20% of the reference method value.
次要结局
未报告次要终点