Respiratory Rate Validation Study - ChroniSense Polso

ChroniSense Medical Ltd.1 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2021年5月10日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
20
试验地点
1
主要终点
Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate

研究概览

简要总结

20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso

详细描述

20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso.

A range of stable respiratory rates were elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.

Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 52 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Subject must have the ability to understand and provide written informed consent
  • Subject is adult over 18 years of age
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker, or a smoker who has refrained from smoking for 1 day.
  • Male or female of any race

排除标准

  • Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
  • Subjects with known respiratory conditions such as:
  • uncontrolled / severe asthma,
  • pneumonia / bronchitis,
  • shortness of breath / respiratory distress,
  • respiratory or lung surgery,
  • emphysema, COPD, lung disease
  • Subjects with self-reported heart or cardiovascular conditions such as:
  • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg
  • have had cardiovascular surgery
  • chest pain (angina)
  • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia
  • previous heart attack
  • blocked artery
  • unexplained shortness of breath
  • congestive heart failure (CHF)
  • history of stroke
  • transient ischemic attack
  • carotid artery disease
  • myocardial ischemia
  • myocardial infarction
  • cardiomyopathy
  • Self-reported health conditions as identified in the Health Assessment Form
  • diabetes,
  • uncontrolled thyroid disease, kidney disease / chronic renal impairment,
  • history of seizures (except childhood febrile seizures),
  • epilepsy,
  • history of unexplained syncope,
  • recent history of frequent migraine headaches,
  • recent head injury within the last 2 months,
  • Cancer / chemotherapy
  • Other known health condition, should be considered upon disclosure in health assessment form

研究组 & 干预措施

ChroniSense Polso Respiratory Rate

Experimental

Comparison of respiratory rates in normal subjects as observational with end tidal CO2. No treatment or interventions will be performed

干预措施: ChoniSense Polso (Device)

结局指标

主要结局

Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate

时间窗: Through study completion, 1 month average

Accuracy of Polso Monitoring System (DUT) relative to a reference End Tidal Carbon Dioxide monitor-derived respiration rate (REF). Polso reports respiration rate (breaths per minute) directly, at 60 second intervals. End Tidal Carbon Dioxide monitor waveforms will be hand-scored, based on respiration waveform peaks, to derive respiration rate (breaths per minute). The readings from each device will be synchronized to achieve paired simultaneous respiration rates, and aggregated into calculation of a comparative accuracy measure, namely Accuracy root-mean-square (Arms) difference, in breaths per minute, between the Polso (DUT) and the reference (REF) respiratory rate system, for all stable respiratory periods. Acceptance criteria Arms \<3.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (1)