Study of Proper Operation of the Syringe Pump Used in Priming of the Syringe
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 48
- 试验地点
- 2
- 主要终点
- delivered infusates of remifentanil
研究概览
简要总结
The investigators attempted to determine an adequate priming volume for our infusion system, and investigated the extent of a possible delay of the drug effect, that would result from mechanical defects of the infusion system, with or without priming the infusion system, using direct gravimetrical measurements of virtual infusate amounts during target controlled infusion of 2 remifentanil diluents.
详细描述
Priming the infusion system (PRIMING) was performed using an evacuation of 2.0 ml to the atmosphere prior to Target-controlled Infusion (TCI). Forty eight TCI, using 50 μg/ml or 20 μg/ml of remifentanil diluents, were performed targeting 4.0 ng/ml of effect site concentration (Ceff), with or without PRIMING. Using simulations, the gravimetrical measurements of the delivered infusates reproduced actual predicted concentrations.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Prevention
- 盲法
- Double (Participant, Investigator)
入排标准
- 年龄范围
- 20 Years 至 70 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •electric medical records of the patients who had undergone general anaesthesia
排除标准
- •body weight exceeding 20% of ideal body weight
结局指标
主要结局
delivered infusates of remifentanil
时间窗: base line from 30 min of TCI was maintained
For targetting an effect-site concentration (Ceff) of 4.0 ng/ml, were randomly performed using 50 μg/ml (Remi50) or 20 μg/ml (Remi20) of remifentanil, and with or without PRIMING, TCI data files, including predicted plasma (Cp-proper), and effect-site (Ceff-proper) concentrations were saved