Effect of CBCT Dose Reduction on the Mandibular Canal Visibility Compared to Standard CBCT Dose Diagnostic Accuracy Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 20
- 试验地点
- 1
- 主要终点
- Primary outcome Evaluation of the diagnostic accuracy of the lowest CBCT dose (Milliampere: 2.2 mA.) compared to the gold standard CBCT dose (Milliampere: 10 mA.) on the mandibular canal visibility.
研究概览
简要总结
The goal of this observational study is to evaluate the diagnostic accuracy of CBCT dose reduction on the mandibular canal visibility, and their effect on canal cortication compared to standard CBCT dose as a gold standard. The main question it aims to answer is: Will the lower CBCT doses affect the visualization of mandibular canal when compared to standard CBCT dose as a gold standard?
研究设计
- 研究类型
- Observational
- 观察模型
- Case Control
- 时间视角
- Prospective
入排标准
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •sound mandibles
排除标准
- •fractured mandibles
研究组 & 干预措施
I4: CBCT imaging with Ultra low dose
exposure parameters:
- Kilovoltage: 90 kVp.
- Milliampere: 5.6 mA.
- Exposure time: 4.5s
- Voxel size: 0.4 mm.
干预措施: CBCT imaging (Diagnostic Test)
I1: Conventional dose CBCT imaging
exposure parameters:
- Kilovoltage: 90 kVp.
- Milliampere: 8 mA.
- Exposure time: 13.5s
- Voxel size: 0.4 mm
干预措施: CBCT imaging (Diagnostic Test)
I2: Conventional dose CBCT imaging
exposure parameters:
- Kilovoltage: 90 kVp.
- Milliampere: 5.6 mA.
- Exposure time: 9 s
- Voxel size: 0.6 mm
干预措施: CBCT imaging (Diagnostic Test)
I3: CBCT imaging with Ultra low dose
exposure parameters:
- Kilovoltage: 90 kVp.
- Milliampere: 7.1 mA.
- Exposure time: 6 s
- Voxel size: 0.2 mm
干预措施: CBCT imaging (Diagnostic Test)
I5: CBCT imaging with Ultra low dose
exposure parameters:
- Kilovoltage: 90 kVp.
- Milliampere: 2.2 mA.
- Exposure time: 4.5 s
- Voxel size: 0.6 mm.
干预措施: CBCT imaging (Diagnostic Test)
R (Reference standard): CBCT imaging with standard dose
exposure parameters:
- Kilovoltage: 90 kVp.
- Milliampere: 10 mA.
- Exposure time: 18s
- Voxel size: 0.2 mm.
干预措施: CBCT imaging (Diagnostic Test)
结局指标
主要结局
Primary outcome Evaluation of the diagnostic accuracy of the lowest CBCT dose (Milliampere: 2.2 mA.) compared to the gold standard CBCT dose (Milliampere: 10 mA.) on the mandibular canal visibility.
时间窗: March 2024
次要结局
未报告次要终点
研究者
Mai Mostafa Mosaad Shaldoum
Researcher
Cairo University