Randomized Sequence, Assessor-Blind, Cross-Over Study of a Windlass Tourniquet, a Pneumatic Tourniquet and an Elastic Adhesive Bandage for Use as a Tourniquet
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 19
- 试验地点
- 2
- 主要终点
- Radiologist Limb Occlusion Assessment
研究概览
简要总结
This cross-over study of twenty five (anticipated) healthy human subjects will utilize two active control devices, and two different widths of the investigational device. One active control is representative of the usual care tourniquet applied in hospital settings. The second active control is a windlass tourniquet representative of the pre-hospital and military usual care tourniquet device. Two inch and four inch widths of the investigational device will be evaluated. The purpose of this study is to characterize and compare investigational and control tourniquet safety and efficacy. All tourniquets will be applied to the upper thigh. Efficacy data is provided by the presence or absence of popliteal flow when assessed with ultrasound. Mapping of interface pressures between the skin and the tourniquet will provide safety data. A minimum washout of five minutes will be allowed between tourniquet applications. All interventions will be applied in one visit. No follow up visit is anticipated.
详细描述
This is a small study of 19 participants conducted over only two days. All data for this study is collected by the study.
Tourniquet application:
All tourniquet applications conducted with the participant seated on the edge of an exam table with one leg hanging off the side, and the leg for tourniquet application extended outward, with the mid calf resting on a support. This position permits access to the popliteal artery for ultrasonography while the investigator has access to apply the tourniquet. The lower edge of each tourniquet is aligned just above the middle of the thigh. All tourniquets are applied over the pressure mapping sensor. The sensor is slightly thicker than denim jeans. Extra care must be taken to ensure that the material of the sensor does not interfere with removing the slack from the windlass tourniquets before turning the windlass to apply final pressure.
When a second windlass tourniquet is to be applied it is applied just above or proximal to the first windlass tourniquet. The windlass of each tourniquet are located as far from each other as possible. Meaning that if the windlass of the first tourniquet is the twelve o'clock position then the windlass of the second tourniquet is positioned at the six o'clock position. The strap of each tourniquet cannot slip over the windlass of the other. This 6 and 12 positioning prevents the strap of the second windlass tourniquet from slipping over the first as it is tightened.
When the tape tourniquet is applied, it is elongated, stretched as it is layered over the preceding layer with at least one half over lap. The tape tourniquet used in this study is similar to, but not the same as the product called manufactured by Entrotech, called "Battle Wrap." Results similar to this study should not be expected with the "Battle Wrap" product.
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Crossover
- 主要目的
- Treatment
- 盲法
- Single (Outcomes Assessor)
入排标准
- 年龄范围
- 18 Years 至 55 Years(Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Upper thigh circumference greater than 61 cm or 24 inches.
- •Upper thigh circumference greater than the 50th percentile based on available anthropometric data.
排除标准
- •Currently taking medications for the treatment of hypertension.
- •Cardiac, renal, pulmonary, hepatic or hemologic disease or disorder.
- •Patients with signs or symptoms of vascular insufficiency. Specifically, patients with any history of non-healing wound, ulcer, blood clots or peripheral vascular disease will be excluded.
- •Any coagulation disorder, prior thrombotic or embolic events such as a deep vein thrombosis.
- •Sickle cell.
- •Diabetes mellitus
- •Rheumatic arthritis or other auto immune disease.
- •Major surgery, significant traumatic injury, within 60 days trial.
- •Skin grafts on lower extremities.
- •Known or observed neurological symptoms, to include peripheral neurological symptoms or deficits.
- •Pregnant women, and women who suspect themselves to be pregnant are not eligible for this study.
- •Any known malignancy.
- •Claudication
研究组 & 干预措施
Pneumatic Tourniquet
All participants randomly allocated to receive all interventions. Pneumatic Tourniquet: The Automatic Tourniquet System (ATS) 1500 by Zimmer (formerly Aspen Labs) is a pneumatic tourniquet that is typically used in surgical settings, and is representative of best outcome under the ideal circumstances of a controlled environment. The 10 cm (4 inch) wide cylindrical cuff was used for all participants.
干预措施: Pneumatic Tourniquet (Device)
Windlass Tourniquet
Windlass Tourniquet with a 3.8 cm (1.5 inch) wide strap is representative of the typical use device in civilian prehospital and military combat environments. In accordance with current prehospital guidelines: If distal perfusion is observed after One Windlass Tourniquet is applied then a second windlass tourniquet will be applied immediately proximal to the first windlass tourniquet so that the participant has Two Windlass Tourniquets applied.
干预措施: One Windlass Tourniquet (Device)
Windlass Tourniquet
Windlass Tourniquet with a 3.8 cm (1.5 inch) wide strap is representative of the typical use device in civilian prehospital and military combat environments. In accordance with current prehospital guidelines: If distal perfusion is observed after One Windlass Tourniquet is applied then a second windlass tourniquet will be applied immediately proximal to the first windlass tourniquet so that the participant has Two Windlass Tourniquets applied.
干预措施: Two Windlass Tourniquets (Device)
Tourniquet Tape 10 cm
Tourniquet Tape 10 cm wide is a highly elastic transparent occlusive tape that is expected to create sufficient circumferential pressure to occlude arterial blood flow when tightly wrapped around a limb. The tape should be applied in mostly overlapping layers. The final wrap should be applied without tension to prevent the elastic tension from causing the tape to unwind itself.
干预措施: Tourniquet Tape 5 cm (Device)
Tourniquet Tape 5 cm
Tourniquet Tape 5 cm wide is a highly elastic transparent occlusive tape that is expected to create sufficient circumferential pressure to occlude arterial blood flow when tightly wrapped around a limb. The tape should be applied in mostly overlapping layers. The final wrap should be applied without tension to prevent the elastic tension from causing the tape to unwind itself.
干预措施: Tourniquet Tape 10 cm (Device)
结局指标
主要结局
Radiologist Limb Occlusion Assessment
时间窗: For 1 sustained minute after application of each tourniquet intervention.
Duplex and color flow ultrasound are used to assess distal blood flow at the popliteal artery for 1 sustained minute following the application of a each tourniquet intervention to the middle upper thigh. If no flow is observed for 1 sustained minute then it is considered successful occlusion. A still image is then saved for later evaluation by the blinded outcome assessor. The 5 cm tape was discontinued after the 4th participant because the tape was considered to unacceptably narrow after stretching and this was considered a mechanical failure even if occlusion was observed. Therefore the 5 cm tape data was not evaluated by the radiologist.
Investigator Limb Occlusion Assessment
时间窗: For 1 sustained minute after application of each tourniquet intervention.
Duplex and color flow ultrasound are used to assess distal blood flow at the popliteal artery for 1 sustained minute following the application of each tourniquet intervention to the middle upper thigh. If no flow is observed for 1 sustained minute then it is considered a successful occlusion. This assessment was made by the investigator and ultrasonographer applying the intervention.
次要结局
未报告次要终点
研究者
Marcus Migura
Sponsor-Investigator
Migura, Marcus