Engineering Evaluation of the Vector NIV Device in Chronic Obstructive Pulmonary Disease (COPD) Patients

Philips Clinical & Medical Affairs Global4 个研究点 分布在 1 个国家目标入组 42 人开始时间: 2018年10月4日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
42
试验地点
4
主要终点
Average EPAP During Overnight PSG

研究概览

简要总结

This study aimed to validate a novel automatic non-invasive ventilation (NIV) mode that continuously adjusts expiratory positive airway pressure (EPAP) to the lowest value that abolishes tidal expiratory flow limitation (EFL). The investigator conducted a prospective, non-randomized, study on stable chronic obstructive pulmonary disease (COPD) patients that may or may not be treated currently with NIV. Patients were studied in a sleep lab on a single night with the auto - titrating EPAP that adjusts to abolish tidal EFL. The primary endpoint was to evaluate the behavior of the EPAP during the night. Additionally, a sub-group of patients used the device at home for a 2 week period. EPAP behavior was assessed during this 2 week period.

详细描述

Tidal Expiratory Flow Limitation (EFLT) is the inability to increase expiratory flow despite increasing effort. Especially in patients with mild to severe COPD, EFL can present challenges when receiving mechanical ventilation. To overcome EFLT expiratory positive airway pressure (EPAP) is applied; however, a single level or fixed EPAP may not overcome the airflow obstruction. The current study was undertaken to explore the variability of EFLT determined using Forced Oscillation Technique (FOT) to dynamically measure lung reactance (∆Xrs) and to evaluate the ability of automatically adjusted EPAP (PEEPopt) to over come EFLT overnight and over a two week period.

In this prospective non-randomized trial, an unreleased noninvasive ventilator set in an S/T mode applied continuous oscillations (5 Hz, 1 cmH2O amplitude, 2 cmH20 peak to peak). Response to the oscillations was analyzed to calculate ∆Xrs and EPAP was adjusted automatically between 4 and 20 cmH2O. A fixed pressure support of 6 cmH2O was delivered. ∆Xrs was measured sitting and supine. Participants with evidence of EFLT (∆Xrs > 2.8) were asked to undergo overnight polysomnography (PSG) and then a sub-group of patients used the ventilator at home for two weeks. EFLT within and between participants was analyzed. EFLT behavior during PSG, the response of the device to dynamically abolish EFLT, as well as the impact of this therapy on sleep quality was studied. Objective ventilator adherence data were used to determine usage after the 2-week in-home use. Additionally, experienced NIV participants qualitatively rated therapy comfort compared to their current treatment.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
40 Years 至 80 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age > 40 years of age; < 80 years of age
  • Ability to provide consent
  • Diagnosis of COPD
  • Must present with EFL via screening of the vector device at 3 cmH2O
  • Have an EPAP to abolish EFL greater or equal to 6cmH2O
  • Must be able to maintain SpO2 greater than 88% at rest and during EPAP Titration
  • Participants who completed the initial study and who would be willing to use the Vector NIV device at home during the night for a 2 week period OR
  • Participants prescribed and currently using a PAP or NIV device at home who meet study inclusion/exclusion criteria of primary protocol

排除标准

  • Any major non COPD uncontrolled disease or condition, such as congestive heart failure, malignancy, liver or renal insufficiency (that requires current evaluation for liver or renal transplantation or dialysis), amyotrophic lateral sclerosis, or severe stroke, or other condition as deemed appropriate by investigator as determined by review of medical history and / or participant reported medical history
  • Suffering from a COPD exacerbation at the time of data collection or in the 30 days prior to data collection
  • Self-reported Pregnancy
  • Employee or family member that is affiliated with Philips Respironics
  • Currently employed by a manufacturer of respiratory products or family member employed by a manufacturer of respiratory products
  • History of bullous emphysema
  • History of pneumothorax
  • Evidence of acute sinusitis or otitis media
  • Hypotension
  • Participants at risk for aspiration of gastric contents
  • Participants in respiratory failure
  • Inability to maintain a patent airway or adequately clear secretions
  • Participants currently using a PAP or NIV device at home with a documented EPAP setting on their current device that is greater than then the mean or final EPAP determined during the therapy session of the screening visit.

结局指标

主要结局

Average EPAP During Overnight PSG

时间窗: Approximately 8 hours

The expiratory positive airway pressure (EPAP) is the parameter that is automatically adjusted. The ventilator automatically titrates EPAP to the minimum value able to abolish expiratory flow limitation.

Difference Between Mean Inspiratory and Expiratory Reactance During Overnight PSG

时间窗: Approximately 8 hours

The difference between mean inspiratory and expiratory reactance is measured using the forced oscillation technique. The difference between mean inspiratory and expiratory reactance is an index of tidal expiratory flow limitation. Values \> 2.8 cmH2O\*s/L are indicative of tidal expiratory flow limitation.

Average EPAP for 2 Week Device Take Home Studies

时间窗: Approximately 8 hours nightly use over 14 days

The expiratory positive airway pressure (EPAP) is the parameter that is automatically adjusted. The ventilator automatically titrates EPAP to the minimum value able to abolish expiratory flow limitation.

Difference Between Mean Inspiratory and Expiatory Reactance During 2 Week Device Take Home Studies

时间窗: 2 weeks

The difference between mean inspiratory and expiatory reactance is measured using the forced oscillation technique. The difference between mean inspiratory and expiratory reactance is an index of tidal expiratory flow limitation. Values \> 2.8 cmH2O\*s/L are indicative of tidal expiratory flow limitation.

次要结局

  • Sleep Quality(Approximately 8 hours)
  • 2 Week Home Use Device Compliance(Nightly use for 14 days)
  • Therapy Comfort Survey(End of 2 weeks device take home study)