临床试验/NCT00804778
NCT00804778
已完成
不适用

Comparison of USCOM Cardiac Output and Continuous Thermodilution Cardiac Output in Patients With Coronary Artery Bypass Grafting

Beijing Chao Yang Hospital1 个研究点 分布在 1 个国家目标入组 18 人2006年3月

概览

阶段
不适用
干预措施
未指定
疾病 / 适应症
Coronary Artery Disease
发起方
Beijing Chao Yang Hospital
入组人数
18
试验地点
1
主要终点
All 18 post-CABG coronary artery disease patients receiving treatment in the SCCU of Beijing Chaoyang Hospital from March to July 2006
状态
已完成
最后更新
17年前

概览

简要总结

It's reported that USCOM can be used to measured patients' CO and CI conveniently, accurately and not confined to place, because of it's no aggressive, it's accuracy is doubtful.

详细描述

18 post-CABG coronary artery disease patients receiving treatment in the SCCU of Beijing Chaoyang Hospital from March to July 2006, male (10) and female (8), with average age of 61.71士10.07 years. Aortic regurgitation or stenosis was ruled out by preoperational color ultrasonic imaging, endocardiac shunt did not exist and there was no heart valve replacement in all the patients. For all the patients after operation, pericardial cavity or left thorax close drainage, small tidal volume (6-8ml/kg) capacity controlled ventilation of breathing machine, appropriate pressure support (10-15cmH2O) and appropriate positive end-expiratory pressure were applied. Intra-aortic balloon pumping (IABP) was not used. Institutional ethics approval was obtained for the conduct of the trial. All the patients received CABG under general anesthesia and non-extracorporeal circulation. Swan-Ganz CCO catheter (mbo 744HF75, USA) was set through right internal jugular vein or right subcalvian vein. The catheters were connected with VIGILANCE monitor after patients went back SCCU and the position of the catheters was adjusted. The bodyweight, height, instant heart rate, central venous pressure, invasive average aortic pressure and pulmonary wedge pressure of the patients were inputted into VIGILANCE monitor. VIGILANCE monitor automatically and intermittently displayed and saved the results of CO, CI, SVR, PVR and other hemodynamic parameters. VIGILANCE determined the hemodynamic parameters every 40-55seconds based on the signal noise ratio. The period of USCOM monitoring time after CABG was 0.5-6hrs. Lowered the bed head and had the patients in horizontal position during the monitoring. Monitoring area was AV area. Turned on USCOM, selected AV mode, inputted the sex, height, bodyweight and birth date of the patients. Applied ultra transmission gel on the probe to let the probe contact the skin closely and let the probe point to the gluteal region nearly. USCOM screen automatically displayed the lateral border of the Doppler blood flow curve scanned with flow monitor and speaker displayed the acoustic signal of blood stream. The position, depth and direction of the probe were adjusted based on the figure signals and acoustic signals acquired with USCOM probe. Adjusted the sound volume, wave gain or contrast or turned the patient head gently. It indicated that the sound wave emitted by the probe had passed through aortic valve orifice or the root of aorta and the direction of sound beam paralleled to the blood flow through the aorta valve orifice when USCOM screen displayed maximum Doppler blood velocity wave and speaker displayed sharp and strong blood acoustic signals. The results were frozen and saved . Optimal measuring site, depth the probe was depressed down and direction of the probe were marked for the next measurement. Continuous 3-5 Doppler flow curves and the CO and CI measured on them were taken each time. The flow curves of one respiratory circle were taken for atrial fibrillation patients. For round probe, plane diameter was 1.5cm and frequency was 3.3MHz. The probe was disinfected with Iodophors each time before and after the measurement. The data of patient height, bodyweight and age, and operation duration was acquired from anesthesia note. CO time intervals measured with the two methods were about 2min. There was at least 30min interval between each pair of CO. Three to four pairs of CO and CI data were taken for each patient.

注册库
clinicaltrials.gov
开始日期
2006年3月
结束日期
2006年7月
最后更新
17年前
研究类型
Observational
性别
All

研究者

发起方
Beijing Chao Yang Hospital

入排标准

入选标准

  • All 18 post-CABG coronary artery disease patients receiving treatment in the SCCU of Beijing Chaoyang Hospital from March to July 2006, male (10) and female (8), with average age of 61.71士10.07 years.

排除标准

  • Aortic regurgitation or stenosis was ruled out by preoperational color ultrasonic imaging, endocardiac shunt did not exist and there was no heart valve replacement in all the patients.
  • Intra-aortic balloon pumping (IABP) was not used.

结局指标

主要结局

All 18 post-CABG coronary artery disease patients receiving treatment in the SCCU of Beijing Chaoyang Hospital from March to July 2006

时间窗: The period of monitoring time after CABG was 0.5-6hrs.

次要结局

  • All 18 post-CABG coronary artery disease patients receiving treatment in the SCCU of Beijing Chaoyang Hospital from March to July 2006(The period of monitoring time after CABG was 0.5-6hrs.)