Entia Liberty: Analytical Performance Validation - Accuracy and Precision

Entia Ltd1 site in 1 country530 target enrollmentStarted: October 3, 2022Last updated:
Conditions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Sponsor
Entia Ltd
Enrollment
530
Locations
1
Primary Endpoint
To obtain the Total Error per parameter from the results of the Entia Liberty device

Study Overview

Brief Summary

This study forms part of Entia's clinical evidence for regulatory submission. This study evaluates Entia Liberty's performance claims with venous blood compared to the gold standard clinical laboratory haematology analysers. Excess blood flagged by the laboratory meeting our requirements will be tested on the Entia Liberty device and have its results compared against the laboratory results. Samples will also be flagged for precision testing (same blood sample, split into 10 Entia Liberty samples).

Study Design

Study Type
Observational
Observational Model
Case Only
Time Perspective
Cross Sectional

Eligibility Criteria

Ages
18 Years to — (Adult, Older Adult)
Sex
All
Accepts Healthy Volunteers
No

Inclusion Criteria

  • Age >18 years old
  • Currently receiving standard of care systemic anti-cancer therapy (chemotherapy, immunotherapy, endocrine and targeted therapy) for solid organ malignancy and has received at least one cycle
  • Blood samples that have been stored in K2-EDTA vacutainers only
  • For Accuracy testing: Samples covering 75% of parameter ranges (Neutrophils: 0.5-7.5 x 109/L, Platelets: 20-450 x 109/L, Haemoglobin: 70-180 g/L), These will be determined by a gold standard analyser which has been recently calibrated.
  • Reliability study: Level range per parameter. These will be determined by a gold standard analyser which has been recently calibrated.
  • Neutrophils (We assume WBC to behave similarly in the same ranges) Low: 0.5-1.2 x 109/L (3 samples), 1.2-2 x 109/L (2 samples) Normal: 2-7 x 109/L (3-4 samples) High: 7-12 x 109/L (1-2 samples)
  • Platelets Low: 50-150 x 109/L (5 samples) Normal: 150-400 x 109/L (3 - 4 samples) High: >400 x 109/L (1-2 samples)
  • Haemoglobin Low: <120 g/L (5 samples) Normal: 120-170 g/L (3 - 4 samples) High: >170 g/L (1-2 samples)

Exclusion Criteria

  • History of haematological malignancy

Outcomes

Primary Outcomes

To obtain the Total Error per parameter from the results of the Entia Liberty device

Time Frame: 3 months

The dataset from this study will provide evidence to support the device's performance claims.

To obtain the precision per parameter from the results of the Entia Liberty device

Time Frame: 3 months

The dataset from this study will provide evidence to support the device's performance claims.

To obtain the bias per parameter from the results of the Entia Liberty device

Time Frame: 3 months

The dataset from this study will provide evidence to support the device's performance claims.

To obtain the linearity per parameter from the results of the Entia Liberty device

Time Frame: 3 months

The dataset from this study will provide evidence to support the device's performance claims.

Secondary Outcomes

No secondary outcomes reported