Clinical Investigation on Assessing the Repeatability and Reproducibility of Central and Peripheral Refractive Error Measurement With the Grand Seiko WAM-5500

Johnson & Johnson Vision Care, Inc.1 个研究点 分布在 1 个国家目标入组 30 人开始时间: 2024年1月16日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
30
试验地点
1
主要终点
Central (on-axis) cycloplegic auto-refraction

研究概览

简要总结

This is a unilateral, non-interventional study. Each subject will be unilaterally measured with the WAM-5500 autorefractor by 3 testers. All measurements will be made on the right eye only.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Potential subjects must satisfy all of the following criteria to be enrolled in the study:
  • Read, understand, and sign the STATEMENT OF INFORMED CONSENT and receive a fully executed copy of the form.
  • Appear able and willing to adhere to the instructions set forth in this clinical protocol.
  • Be between 18 and 50 (inclusive) years of age at the time of screening.
  • The spherical component of the subject's non-vertex-corrected distance subjective refraction must be between +1.50 and -6.00DS (inclusive), with the magnitude of the cylinder 2.00 DC or less in the right eye.
  • The subject must have a Van Herrick angle grade of at least 3 or greater in the right eye.
  • The subject must have a BSCVA of at least 20/30 (defined as reading at least 3 out of the 5 letters on the 20/30 line) in the right eye.

排除标准

  • Potential subjects who meet any of the following criteria will be excluded from participating in the study:
  • Be currently pregnant or lactating.
  • Be diabetic or pre-diabetic by self-report.
  • Have a history of ocular trauma, systemic disease or medication use known to cause variability in refractive error.
  • Have had or have planned (within the study period) any ocular or intraocular surgery (e.g., radial keratotomy, PRK, LASIK, iridotomy, retinal laser photocoagulation, etc.).
  • Have habitually worn rigid gas permeable (RGP) lenses, orthokeratology lenses, scleral lenses, or hybrid lenses (e.g. SynergEyes, SoftPerm) within the past 3 months.
  • Be an employee (e.g., Investigator, Coordinator, Technician) or immediate family member of an employee (including partner, child, parent, grandparent, grandchild or sibling of the employee or their spouse) of the clinical site.
  • Take ocular and systemic medications known to interact with Tropicamide ophthalmic solution or systemic medications that can cause myopia.
  • Have a history of an allergic reaction to Tropicamide or Proparacaine.
  • Have clinically significant (grade 3 or higher on the FDA grading scale) slit lamp findings (e.g., corneal edema, neovascularization or staining, tarsal abnormalities, or bulbar injection) or other corneal or ocular disease or abnormalities that may compromise study endpoints (including entropion, ectropion, chalazia, recurrent styes, glaucoma, corneal scar or opacity ≥ 0.3 mm, history of recurrent corneal erosions, aphakia, moderate or above corneal distortion, herpetic keratitis).
  • Have fluctuations in vision due to clinically significant dry eye or other ocular conditions.
  • Have a history of strabismus, amblyopia, nystagmus, or any condition that affects fixation.

结局指标

主要结局

Central (on-axis) cycloplegic auto-refraction

时间窗: Up to 10-day follow-up visit

Central (on-axis) cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of central refractive error measurements with the Grand Seiko WAM-5500.

Peripheral (20° off-axis on the temporal retina) cycloplegic auto-refraction

时间窗: Up to 10-day follow-up visit

Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 20° temporal retina.

Peripheral (10° off-axis on the temporal retina) cycloplegic auto-refraction

时间窗: Up to 10-day follow-up visit

Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 10° temporal retina.

Peripheral (10° off-axis on the nasal retina) cycloplegic auto-refraction

时间窗: Up to 10-day follow-up visit

Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 10° nasal retina.

Peripheral (20° off-axis on the nasal retina) cycloplegic auto-refraction:

时间窗: Up to 10-day follow-up visit

Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 20° nasal retina.

Peripheral (30° off-axis on the temporal retina) cycloplegic auto-refraction

时间窗: Up to 10-day follow-up visit

Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 30° temporal retina.

Peripheral (30° off-axis on the nasal retina) cycloplegic auto-refraction

时间窗: Up to 10-day follow-up visit

Peripheral cycloplegic auto-refraction will be measured to evaluate the intra-visit repeatability, inter-visit repeatability, and reproducibility of peripheral refractive error measurements with the Grand Seiko WAM-5500 at 30° nasal retina.

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...

相似试验