Accuracy of ScanWatch Pulse Oximeter With Profound Hypoxia
- Conditions
- Oximetry
- Interventions
- Diagnostic Test: hypoxia
- Registration Number
- NCT04380389
- Lead Sponsor
- Withings
- Brief Summary
Performance testing of pulse oximeters. The protocol involves brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample is analyzed for oxygen saturation with a gold standard bench CO-oximeter. This protocol is aligned with the latest ISO and FDA guidance documents for pulse oximeter testing.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 14
- The subject is male or female, aged ≥18 and <50.
- The subject is in good general health with no evidence of any medical problems.
- The subject has both wrist circumferences between 14 and 22cm.
- The subject has provided informed consent and is willing to comply with the study procedures.
- The subject is obese (BMI>30).
- The subject has a known history of heart disease, lung disease, kidney or liver disease.
- Diagnosis of asthma, sleep apnea, or use of CPAP.
- Subject has diabetes.
- Subject has a clotting disorder.
- The subject a hemoglobinopathy or history of anemia, per subject report or the first blood sample that in the opinion of the investigator, would make them unsuitable for study participation.
- The subject has any other serious systemic illness.
- The subject is a current smoker.
- The subject has piercings that may cause air leaks during the test
- The subject has any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.
- The subject has a history of fainting or vasovagal response.
- The subject has a history of sensitivity to local anesthesia.
- The subject has a diagnosis of Raynaud's disease.
- The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
- The subject is pregnant, lactating or trying to get pregnant.
- The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
- The subject has a resting heartrate over 120, Systolic BP is over 150, diastolic BP is over 90 or room air SpO2 is less than 94%.
- The subject has carboxyhemoglobin over 3% as measured by Lab blood gas analysis.
- The subject has any other condition, which in the opinion of the investigators' would make them unsuitable to participate.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description profound hypoxia hypoxia -
- Primary Outcome Measures
Name Time Method RMSE compared to CO-oximetry 1 hour Root Mean Squared Error between the functional blood oxygen saturation (SpO2) provided by the device under test and the arterial blood oxygen measured with the CO-oximeter.
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Hypoxia Lab, UCSF
🇺🇸San Francisco, California, United States