Accuracy of the Dexcom G6 Continuous Glucose Monitoring System Following Cardiac Surgery

The Cleveland Clinic1 个研究点 分布在 1 个国家目标入组 40 人开始时间: 2021年5月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
40
试验地点
1
主要终点
Number of Matching Glucose Measurements (CGM Versus Blood Glucose Measurements)

研究概览

简要总结

This is a prospective longitudinal study to study the agreement between a novel continuous glucose monitoring system (CGMS) versus current blood glucose monitoring.

Subjects in this study will have their blood glucose measured regularly every 1-3 hours with current methods in the cardiovascular intensive care unit (CVICU), and by point of care glucose using the Accuchek Inform II on the regular floors, and the CGMS at the same time will be captured. Subjects will have measurements taken throughout their stay in the CVICU and on the regular floors.

详细描述

Maintaining tight blood glucose control following surgery is imperative to reducing infections and neurologic dysfunction. This requires frequent blood sampling while in the intensive care unit, leading to increased waste of blood and utilizing time and resources to collect samples and wait for results. Additionally, when on the regular floors, frequent point of care fingersticks for glucose levels are needed for therapeutic intervention.

Continuous glucose monitoring systems (CGMS) are now available for ambulatory use in patients with diabetes. The potential of using such a system in the hospital, including during the postoperative critical care setting and upon transfer to the regular floor, may reduce personnel burden, produce rapid results, and decrease blood waste. However, these systems have not yet been validated for hospital use.

This is a prospective longitudinal study to study the agreement between a novel continuous glucose monitoring system (CGMS) versus current blood glucose monitoring.

Subjects in this study will have their blood glucose measured regularly every 1-3 hours with current methods in the CVICU, and by point of care glucose using the Accuchek Inform II on the regular floors, and the CGMS at the same time will be captured. Subjects will have measurements taken throughout their stay in the CVICU and on the regular floors. Agreement and correlation between systems will be calculated, and error will be classified using the Surveillance Error grid. The proportion of errors of at least moderate risk will be calculated.

Subject characteristics including demographics, comorbidities and baseline disease severity will be summarized using means and standard deviations for continuous measures and frequencies and percentages for categorical factors. To evaluate agreement between methods, concordance correlation coefficients will be calculated, using the method that adjusts for repeated measures as described by King and implemented in software by Carrasco. Analysis will be performed overall, and then stratifying by type of current measure, if the type of measurement varies across settings. Correlations will be calculated using the methods described by Bland and Altman and implemented by Bakdash and Marusich. Differences will also be evaluated for errors according to the Surveillance Error Grid. Rates of moderate or more severe errors will be calculated. For each measure above, 95% confidence intervals will be calculated, accounting for the clustering within subject.. Analyses will be performed using SAS software (version 9.4; Cary, NC) and R software (version 3.5; Vienna Austria).

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Supportive Care
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age 18 years and above
  • Planned cardiovascular surgery
  • Planned admission to Cleveland Clinic Main Campus J5 or J6 or Q5 cardiovascular ICU (CVICU) post-operatively
  • With or without known diabetes; if with known diagnosis of diabetes, diabetes can be type 1, type 2, or secondary

排除标准

  • Allergy to the material of the CGMS or the adhesive to be used
  • Skin conditions precluding the use of the CGMS
  • Pregnancy
  • Other conditions that investigators deem inappropriate for the study

研究组 & 干预措施

Dexcom G6 continuous glucose monitor

Other

All patients will have a Dexcom G6 continuous glucose monitor placed pre-operatively. The glucose readings will be collected for 10 days or upon discharge from the ICU and data will be compared with arterial blood glucose readings or venous Accu-Check Inform II glucose readings

干预措施: Dexcom G6 continuous glucose monitor (Device)

结局指标

主要结局

Number of Matching Glucose Measurements (CGM Versus Blood Glucose Measurements)

时间窗: Up to 10 days

Glucose readings from the Dexcom continuous glucose monitor compared with blood glucose from the arterial blood gas (ABG) or peripheral/central venous catheter

Glucose Readings From Continuous Glucose Monitor (CGM) Versus Point of Care Glucose

时间窗: up to 10 days