Evaluation of Blood Glucose Meter Systems

Not Applicable
Completed
Conditions
Diabetes
Interventions
Device: Contour® PLUS BGMS
Device: OneTouch® SelectSimple™ BGMS
Device: Accu-Chek® Performa BGMS
Device: Accu-Chek® Active BGMS
Device: Freestyle Freedom® BGMS
Registration Number
NCT01714232
Lead Sponsor
Ascensia Diabetes Care
Brief Summary

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Detailed Description

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems were evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
106
Inclusion Criteria
  • Males and females, 18 years of age or older
  • Willing to complete all study procedures
Exclusion Criteria
  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Pregnancy

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Study Staff Test BGMSsContour® PLUS BGMSAll testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
Study Staff Test BGMSsAccu-Chek® Performa BGMSAll testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
Study Staff Test BGMSsAccu-Chek® Active BGMSAll testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
Study Staff Test BGMSsFreestyle Freedom® BGMSAll testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
Study Staff Test BGMSsOneTouch® SelectSimple™ BGMSAll testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
Primary Outcome Measures
NameTimeMethod
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range8 hours

Using the overall Blood Glucose (BG) range (27 to 460 mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.

Secondary Outcome Measures
NameTimeMethod
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range(<=80 mg/dL)8 hours

Using fresh and glycolyzed samples with Blood Glucose (BG) \<=80 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.

MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)8 hours

Using samples with Blood Glucose \>180 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.

Trial Locations

Locations (1)

Bayer HealthCare LLC, Diabetes Care

🇺🇸

Mishawaka, Indiana, United States