Evaluation of Blood Glucose Meter Systems
- Conditions
- Diabetes
- Interventions
- Device: Contour® PLUS BGMSDevice: OneTouch® SelectSimple™ BGMSDevice: Accu-Chek® Performa BGMSDevice: Accu-Chek® Active BGMSDevice: Freestyle Freedom® BGMS
- Registration Number
- NCT01714232
- Lead Sponsor
- Ascensia Diabetes Care
- Brief Summary
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
- Detailed Description
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems were evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 106
- Males and females, 18 years of age or older
- Willing to complete all study procedures
- Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
- Hemophilia or any other bleeding disorder
- Pregnancy
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Study Staff Test BGMSs Contour® PLUS BGMS All testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS. Study Staff Test BGMSs Accu-Chek® Performa BGMS All testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS. Study Staff Test BGMSs Accu-Chek® Active BGMS All testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS. Study Staff Test BGMSs Freestyle Freedom® BGMS All testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS. Study Staff Test BGMSs OneTouch® SelectSimple™ BGMS All testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
- Primary Outcome Measures
Name Time Method MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range 8 hours Using the overall Blood Glucose (BG) range (27 to 460 mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.
- Secondary Outcome Measures
Name Time Method MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range(<=80 mg/dL) 8 hours Using fresh and glycolyzed samples with Blood Glucose (BG) \<=80 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL) 8 hours Using samples with Blood Glucose \>180 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.
Trial Locations
- Locations (1)
Bayer HealthCare LLC, Diabetes Care
🇺🇸Mishawaka, Indiana, United States