Comparison Evaluation of Blood Glucose Meter Systems

Not Applicable
Completed
Conditions
Diabetes
Interventions
Device: CONTOUR® NEXT EZ BGMS
Device: FreeStyle Freedom Lite® BGMS
Device: ACCU-CHEK® Aviva BGMS
Device: TRUEtrack® BGMS
Device: OneTouch® Ultra®2 BGMS
Registration Number
NCT01548885
Lead Sponsor
Ascensia Diabetes Care
Brief Summary

The purpose of this study was to evaluate the performance of one investigational Bayer Blood Glucose meter (with investigational Bayer test strips) and four Blood Glucose Monitoring Systems from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems was evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some blood samples were glycolyzed to lower the glucose levels.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
146
Inclusion Criteria
  • Males and females
  • 18 years of age or older
  • Willing to complete all study procedures
Exclusion Criteria
  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Pregnancy

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Study Staff Test BGMSsACCU-CHEK® Aviva BGMSAll testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Study Staff Test BGMSsCONTOUR® NEXT EZ BGMSAll testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Study Staff Test BGMSsOneTouch® Ultra®2 BGMSAll testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Study Staff Test BGMSsFreeStyle Freedom Lite® BGMSAll testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Study Staff Test BGMSsTRUEtrack® BGMSAll testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Primary Outcome Measures
NameTimeMethod
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range10 hours

Using the overall Blood Glucose(BG) range (24 to 386mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.

Secondary Outcome Measures
NameTimeMethod
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)10 hours

Using fresh and glycolyzed samples with Blood Glucose(BG) below 70 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.

Trial Locations

Locations (1)

Bayer HealthCare LLC, Diabetes Care

🇺🇸

Mishawaka, Indiana, United States