Comparison Evaluation of Blood Glucose Meter Systems
- Conditions
- Diabetes
- Interventions
- Device: CONTOUR® NEXT EZ BGMSDevice: FreeStyle Freedom Lite® BGMSDevice: ACCU-CHEK® Aviva BGMSDevice: TRUEtrack® BGMSDevice: OneTouch® Ultra®2 BGMS
- Registration Number
- NCT01548885
- Lead Sponsor
- Ascensia Diabetes Care
- Brief Summary
The purpose of this study was to evaluate the performance of one investigational Bayer Blood Glucose meter (with investigational Bayer test strips) and four Blood Glucose Monitoring Systems from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems was evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some blood samples were glycolyzed to lower the glucose levels.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 146
- Males and females
- 18 years of age or older
- Willing to complete all study procedures
- Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
- Hemophilia or any other bleeding disorder
- Pregnancy
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Study Staff Test BGMSs ACCU-CHEK® Aviva BGMS All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS. Study Staff Test BGMSs CONTOUR® NEXT EZ BGMS All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS. Study Staff Test BGMSs OneTouch® Ultra®2 BGMS All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS. Study Staff Test BGMSs FreeStyle Freedom Lite® BGMS All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS. Study Staff Test BGMSs TRUEtrack® BGMS All testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
- Primary Outcome Measures
Name Time Method MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range 10 hours Using the overall Blood Glucose(BG) range (24 to 386mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
- Secondary Outcome Measures
Name Time Method MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL) 10 hours Using fresh and glycolyzed samples with Blood Glucose(BG) below 70 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.
Trial Locations
- Locations (1)
Bayer HealthCare LLC, Diabetes Care
🇺🇸Mishawaka, Indiana, United States